background image

Hazardous Voltage

 This icon denotes a warning advising you to take

precautions to avoid electrical shock.

Safety Guidelines for Hazardous Voltages

If hazardous voltages are connected to the device, take the following precautions. A

hazardous voltage is a voltage greater than 42.4 Vpk voltage or 60 VDC to earth

ground.

Caution

 Ensure that hazardous voltage wiring is performed only by

qualified personnel adhering to local electrical standards.

Caution

 Do not mix hazardous voltage circuits and human-accessible

circuits on the same module.

Caution

 Ensure that devices and circuits connected to the module are

properly insulated from human contact.

Caution

 When module terminals are hazardous voltage LIVE (>42.4 Vpk/

60 VDC), you must ensure that devices and circuits connected to the

module are properly insulated from human contact. You must use the

TB-9212 included with the NI 9212 to ensure that the terminals are not

accessible.

Note

 The TB-9212 with screw terminal contains a plastic insert to prevent

accidental wire contact with the metal enclosure.

Isolation Voltages

NI 9212 and TB-9212 with Screw Terminal Isolation Voltages

Connect only voltages that are within the following limits:

Channel-to-channel isolation

ni.com

4

NI-9212

Содержание NI 9212

Страница 1: ...NI 9212 2023 06 07...

Страница 2: ...arine Applications 8 Preparing the Environment 9 TB 9212 Pinout 9 NI 9212 Connection Guidelines 10 Minimizing Thermal Gradients 10 NI 9212 and TB 9212 with Screw Terminal Thermocouple Connection 11 NI...

Страница 3: ...nt in the system to determine the safety and EMC ratings for the entire system 2015 2016 National Instruments Corporation All rights reserved Refer to the National Instruments _Legal Information direc...

Страница 4: ...ts on the same module Caution Ensure that devices and circuits connected to the module are properly insulated from human contact Caution When module terminals are hazardous voltage LIVE 42 4 Vpk 60 VD...

Страница 5: ...ion system referred to as MAINS voltage MAINS is a hazardous live electrical supply system that powers equipment This category is for measurements of voltages from specially protected secondary circui...

Страница 6: ...tinuous 60 VDC Measurement Category I Withstand 1 000 Vrms Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as...

Страница 7: ...azardous Caution Do not remove modules unless power has been switched off or the area is known to be nonhazardous Caution Substitution of components may impair suitability for Class I Division 2 Cauti...

Страница 8: ...ith the regulatory requirements and limits for electromagnetic compatibility EMC stated in the product specifications These requirements and limits provide reasonable protection against harmful interf...

Страница 9: ...r filtered power and input output ports In addition take precautions when designing selecting and installing measurement probes and cables to ensure that the desired EMC performance is attained Prepar...

Страница 10: ...conducting heat directly to terminal junctions can cause thermal gradients Observe the following guidelines to minimize thermal gradients and improve the system accuracy Use small gauge thermocouple...

Страница 11: ...ermocouple Connection Thermocouple TC TC Shield TB 9212 NI 9212 Ground Lug NI 9212 and TB 9212 with Mini TC Thermocouple Connection Caution Electrostatic Discharge ESD can damage the TB 9212 with mini...

Страница 12: ...ximum torque of 0 4 N m 3 6 lb in Do not overtighten the jackscrews Wiring the TB 9212 with screw terminal What to Use TB 9212 with screw terminal 0 05 mm to 0 5 mm 30 AWG to 20 AWG wire with 5 1 mm 0...

Страница 13: ...no exposed wire extends past the screw terminal 3 Route the wire through the TB 9212 with screw terminal opening remove slack from the wiring and secure the wires using a zip tie 4 Replace the foam pa...

Страница 14: ...orque of 0 4 N m 3 6 lb in Do not overtighten the jackscrews Connecting the TB 9212 with mini TC What to Use TB 9212 with mini TC Shielded thermocouple Clamp on ferrite bead part number 781233 01 What...

Страница 15: ...t self help resources to email and phone assistance from NI Application Engineers Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com r...

Страница 16: ...es create your service request at ni com support or dial 1 866 ASK MYNI 275 6964 For telephone support outside the United States visit the Worldwide Offices section of ni com niglobal to access the br...

Отзывы: