NI 9209 Calibration Procedure
|
© National Instruments
|
5
b.
Configure the AI voltage task according to Table 3. Set the ADC timing mode to
high-resolution.
c.
Start the task.
d.
Average the readings.
e.
Stop and clear the task.
6.
Compare the average to the limits in Table 4.
Note
The test limits listed in Table 4 are derived using the values in Table 8.
Note
The uncertainty of the calibration standard is not included in these limits.
7.
Repeat step 5 and 6 with the ADC timing mode on the NI 9209 set to high-speed.
8.
Set the calibrator to Standby mode (STBY).
9.
Repeat steps 3 through 8 for each test point listed in Table 4.
10. Disconnect the calibrator from the device.
11. Repeat steps 1 through 10 for each voltage channel on the NI 9209.
Table 3.
NI 9209 Voltage Timing Configuration
ADC Timing
Mode
Sample Mode
Samples to Read
Sample Rate
High-Resolution
N Samples
5
19 S/s
High-Speed
N Samples
75
500 S/s
Table 4.
NI 9209 Verification Test Limits for Differential Voltage Accuracy
Range (V)
Test Point
2-Year Limits
Minimum
Maximum
Location
Value
(V)
Lower Limit
(V)
Upper Limit
(V)
-10.00000
10.00000
Pos FS
10.00000
9.98972
10.01028
-10.00000
10.00000
Mid
0.00000
-0.00058
0.00058
-10.00000
10.00000
Neg FS
-10.00000
-10.01028
-9.98972