NI 9209 Calibration Procedure

|

© National Instruments

|

5

b.

Configure the AI voltage task according to Table 3. Set the ADC timing mode to 
high-resolution.

c.

Start the task.

d.

Average the readings.

e.

Stop and clear the task.

6.

Compare the average to the limits in Table 4.

Note

The test limits listed in Table 4 are derived using the values in Table 8.

Note

The uncertainty of the calibration standard is not included in these limits.

7.

Repeat step 5 and 6 with the ADC timing mode on the NI 9209 set to high-speed.

8.

Set the calibrator to Standby mode (STBY).

9.

Repeat steps through 8 for each test point listed in Table 4.

10. Disconnect the calibrator from the device.
11. Repeat steps through 10 for each voltage channel on the NI 9209. 

Table 3.  

NI 9209 Voltage Timing Configuration

ADC Timing 

Mode

Sample Mode

Samples to Read

Sample Rate

High-Resolution

N Samples

5

19 S/s

High-Speed

N Samples

75

500 S/s

Table 4.  

NI 9209 Verification Test Limits for Differential Voltage Accuracy

Range (V)

Test Point

2-Year Limits

Minimum

Maximum

Location

Value 

(V)

Lower Limit 

(V)

Upper Limit 

(V)

-10.00000

10.00000

Pos FS

10.00000

9.98972

10.01028

-10.00000

10.00000

Mid

0.00000

-0.00058

0.00058

-10.00000

10.00000

Neg FS

-10.00000

-10.01028

-9.98972

Содержание NI 9209

Страница 1: ...rential Voltage Accuracy Verification 4 Single Ended Voltage Accuracy Verification 6 Adjustments 7 Voltage Accuracy Adjustment 7 EEPROM Update 8 Reverification 8 Accuracy Under Calibration Conditions...

Страница 2: ...NI 9209 specific information NI 9209 Datasheet NI 9209 specifications and calibration interval NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW Help LabVIEW...

Страница 3: ...iminate noise and thermal offsets Maintain an ambient temperature of 23 5 C The device temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of...

Страница 4: ...librator as shown in Figure 1 Figure 1 Differential Voltage Accuracy Connections Note If the calibrator outputs are truly floating connect the negative output to a quiet earth ground as well as COM to...

Страница 5: ...Set the calibrator to Standby mode STBY 9 Repeat steps 3 through 8 for each test point listed in Table 4 10 Disconnect the calibrator from the device 11 Repeat steps 1 through 10 for each voltage chan...

Страница 6: ...e to a Test Point value indicated in Table 6 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage task on the NI 9209 according to Table 5 b Conf...

Страница 7: ...red to adjust the NI 9209 Voltage Accuracy Adjustment Complete the following procedure to adjust the voltage accuracy performance of the NI 9209 1 Set the calibrator to Standby mode STBY 2 Connect the...

Страница 8: ...de the physical channel s should be a union of the channels specified in both configurations n Close and commit the calibration session 4 Remove the connections from the NI 9209 EEPROM Update When an...

Страница 9: ...rs Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your National Instruments product Product registration faci...

Страница 10: ...nts You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal exp...

Отзывы: