6
|
ni.com
|
NI 9207 Calibration Procedure
5.
Acquire and average samples.
a.
Create and configure an AI voltage task on the NI 9207 according to Table 5.
b.
Configure the AI current task according to Table 6. Set the ADC timing mode to
high-resolution.
c.
Start the task.
d.
Average the readings.
e.
Clear the task.
6.
Compare the average to the limits in Table 7.
7.
Repeat steps 5 and 6 with the ADC timing mode on the NI 9207 set to high-speed.
8.
Set the calibrator to Standby mode (STBY).
9.
Repeat steps 3 through 8 for each test point in Table 7.
10. Disconnect the calibrator from the device.
11. Repeat steps 1 through 10 for each current channel on the NI 9207.
Table 5.
NI 9207 Configuration for Current Accuracy Verification
Input Range
Scaled Units
Terminal Configuration
Min
Max
-0.02
0.02
Amps
RSE
Table 6.
Current Timing Configuration
ADC Timing Mode
Sample Mode
Samples to Read
High-Resolution
N Samples
5
High-Speed
N Samples
75
Table 7.
NI 9207 Test Limits for Current Accuracy Verification
Range (A)
Test Point
1-Year Limits
Minimum
Maximum
Location
Value (A)
Lower Limit (A)
Upper Limit (A)
-0.020000
0.020000
Pos FS
0.021000
0.020970
0.021030
-0.020000
0.020000
Mid
0.000000
-0.000002
0.000002
-0.020000
0.020000
Neg FS
-0.021000
-0.021030
-0.020970