background image

©

 National Instruments Corporation

11

NI 5922 Calibration Procedure

Allow a warm-up time of at least 15 minutes after NI-SCOPE loads. 
Unless manually disabled, NI-SCOPE automatically loads with the 
operating system and enables the device. The warm-up time ensures 
that the measurement circuitry of the NI 5922 is at a stable operating 
temperature.

Ensure that the PXI chassis fan speed is set to HI, that the fan filters are 
clean, and that the empty slots contain filler panels.

Plug the PXI chassis and the calibrator into the same power strip to 
avoid ground loops.

External Calibration Procedures

The complete external calibration procedure consists of self-calibrating, 
verifying the performance of the digitizer, adjusting the calibration 
constants, and verifying again after the adjustments. In some cases, the 
complete calibration procedure may not be required. Refer to the 

External 

Calibration Options

 section for more information.

The external calibration procedure automatically stores the calibration date 
to allow traceability.

Verifying NI 5922 Specifications

Note

After the 15-minute warm-up time, always self-calibrate the NI 5922 before 

beginning a verification procedure.

This section describes the program you must write to verify either the 
calibration test limits or the published specifications for the NI 5922. Refer 
to the 

External Calibration Options

 section to determine which limits to 

use in these procedures.

All verification procedures described in this section begin with 

niScope_init

 (niScope Initialize VI) with 

resetDevice

 set to 

NISCOPE_VAL_TRUE

 and end with 

niScope_close

 (niScope Close VI).

If any of these tests fail immediately after you perform an external 
adjustment, verify that you have met the required test conditions listed in 
the 

External Calibration Requirements

 section before you return the 

digitizer to NI for repair.

Содержание NI 5922

Страница 1: ...on 2 Calibration Intervals 2 Software and Documentation 2 Self Calibration Procedures 4 MAX 5 Scope SFP 5 NI SCOPE 5 External Calibration Options 6 Complete Calibration 7 Optional Calibration 8 Extern...

Страница 2: ...constants in the EEPROM and is equivalent to a factory calibration at NI Because the external calibration procedure changes all EEPROM constants it invalidates the original calibration certificate If...

Страница 3: ...nding LabVIEW VI in parentheses The C function calls are valid for any compiler capable of calling a 32 bit DLL Many of the functions use constants defined in the niScopeCal h file To use these consta...

Страница 4: ...ing for any error measured in the internal references Note Self calibrate the digitizer before you perform verification NI SCOPE includes self calibration example programs for LabVIEW CVI and Microsof...

Страница 5: ...om the Scope SFP complete the following steps 1 Disconnect or disable any AC inputs to the digitizer 2 Launch the Scope SFP by selecting Start Programs National Instruments NI SCOPE Scope Soft Front P...

Страница 6: ...on can be used to ensure that the adjustment process was successful or to determine if the adjustment process must be performed During verification you must compare the measurement error to limits giv...

Страница 7: ...orming a complete calibration is the recommended way to guarantee that the NI 5922 meets or exceeds its published specifications for a two year calibration interval At the end of the complete calibrat...

Страница 8: ...ion date effectively resetting the calibration interval without actually performing an adjustment Refer to the Adjusting the NI 5922 section for more information If all of the measurement errors deter...

Страница 9: ...mits Yes No Meets Published Specifications Yes Results Within Published Specs Adjustment Optional Update Calibration Dates and Temperatures Without Adjusting No Adjust Cal Dates and Temperatures Updat...

Страница 10: ...brator to the input BNC of the digitizer or use a short 50 BNC coaxial cable if necessary Long cables and wires act as antennae picking up extra noise that can affect measurements Keep relative humidi...

Страница 11: ...ion procedure may not be required Refer to the External Calibration Options section for more information The external calibration procedure automatically stores the calibration date to allow traceabil...

Страница 12: ...ConfigureVertical niScope Configure Vertical VI with the following parameters channelList 0 range The Range value for the current iteration from Table 3 offset 0 0 coupling NISCOPE_VAL_DC probeAttenua...

Страница 13: ...the BNC inputs of the NI 5922 No input signal is needed because the NI 5922 internal 50 resistor connects the input to ground 2 Call niScope_ConfigureChanCharacteristics niScope Configure Chan Charac...

Страница 14: ...ned from Table 4 a Add up the squares of each element in the returned array starting with element i and ending with element 26 215 The first element in the array is element 0 b Take the square root of...

Страница 15: ...cal niScope Configure Vertical VI with the following parameters channelList 0 range 2 offset 0 0 coupling NISCOPE_VAL_DC probeAttenuation 1 0 enabled NISCOPE_VAL_TRUE 4 Call niScope_ConfigureHorizonta...

Страница 16: ...maxInputFrequency 0 8 Call niScope_InitiateAcquisition niScope Initiate Acquisition VI with the following parameter vi The instrument handle you obtained from niScope_init 9 Call niScope_FetchMeasure...

Страница 17: ...the following parameters channelList 0 range The Range value for the current iteration from Table 6 offset 0 0 coupling NISCOPE_VAL_DC probeAttenuation 1 0 enabled NISCOPE_VAL_TRUE 3 Call niScope_Con...

Страница 18: ...etchMeasurement niScope Fetch Measurement VI with the following parameters channelList 0 timeout 1 0 scalarMeasFunction NISCOPE_VAL_VOLTAGE_AVERAGE 13 Calculate the error in the vertical gain as a per...

Страница 19: ...ical niScope Configure Vertical VI with the following parameters channelList 0 range 10 offset 0 0 coupling NISCOPE_VAL_DC probeAttenuation 1 0 enabled NISCOPE_VAL_TRUE 3 Call niScope_ConfigureHorizon...

Страница 20: ...nction NISCOPE_VAL_FFT_FREQUENCY 9 Calculate the error in timing as parts per million ppm using the following formula where a the measured frequency Compare the result to the Calibration Test Limits o...

Страница 21: ...ltime NISCOPE_VAL_TRUE 4 Call niScope_ConfigureTriggerEdge niScope Configure Trigger Edge VI with the following parameters triggerSource 0 level 0 1 slope NISCOPE_VAL_POSITIVE triggerCoupling NISCOPE_...

Страница 22: ...ements within array X2 so that index 0 of the array contains the element with the largest value Denote the result as X3 A circular rotation takes the first element of the array makes it the last eleme...

Страница 23: ...listed in the Calibration Test Limits column or the Published Specifications column in Table 8 If the result is less than the selected test limit the device has passed this portion of the verification...

Страница 24: ...0 0 coupling NISCOPE_VAL_DC probeAttenuation 1 0 enabled NISCOPE_VAL_TRUE 3 Call niScope_ConfigureHorizontalTiming niScope Configure Horizontal Timing VI with the following parameters minSampleRate 1...

Страница 25: ...ized data from the input channel ADC when performing an analog trigger on CH 0 or CH 1 Verifying the trigger sensitivity for the input channels is not necessary Adjusting the NI 5922 If the NI 5922 su...

Страница 26: ...SelfCalibrate niScope Cal Self Calibrate VI with the following parameters channelList VI_NULL option VI_NULL 6 Call niScope_CalAdjustVCXO niScope Cal Adjust VCXO VI with the following parameter stimul...

Страница 27: ...you can retrieve all the data programmatically using NI SCOPE functions MAX To retrieve data using MAX complete the following steps 1 Select the device from which you want to retrieve information from...

Страница 28: ...ct supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin...

Отзывы: