NI 5421/5441 Calibration Procedure
36
ni.com
5.
Call
niFgen_SetAttributeViBoolean
to set the analog filter state
(niFgen property node:
Output Attributes»Analog Filter Enabled
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ANALOG_FILTER_ENABLED
•
value
: The
Analog Filter Enable
value for the current iteration
from Table 7
6.
Call
niFgen_SetAttributeViReal64
to set the pre-amplifier
attenuation (niFgen property node:
Calibration»Pre-Amplifier
Attenuation
) using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_PRE_AMPLIFIER_ATTENUATION
•
value
: The
Pre-Amplifier Attenuation
value for the current
iteration from Table 7
7.
Take the following voltage measurements at the NI 5421/5441 CH 0
front panel connector into a high-impedance load:
a.
Call
niFgen_SetAttributeViInt32
to set the gain DAC value
(niFgen property node:
Calibration»Gain DAC Value
) using the
following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_GAIN_DAC_VALUE
•
value
:
2000
b.
Call
niFgen_SetAttributeViInt32
to set the offset DAC
value (niFgen property node:
Calibration»Offset DAC Value
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_InitExtCal
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_OFFSET_DAC_VALUE
•
value
:
50000
c.
Call
niFgen_Commit
(niFgen Commit VI) to commit the
attribute values to the device using the following parameter:
•
vi
: The session handle returned from
niFgen_InitExtCal