National Instruments Corporation
31
NI 5402/5406 Calibration Procedure
4.
Set the output impedance by calling the niFgen Property Node and
selecting
Basic Operation»Output Impedance
.
Repeat steps 5 through 13 for each of the iterations listed in Table 6,
changing the Analog Static Value, Amplitude, and DC Offset values for
each iteration.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_init
.
channelName
:
"0"
attributeID
:
NIFGEN_ATTR_OUTPUT_
IMPEDANCE
value
:
50.00
Table 6.
Values for Verifying Square Wave Gain and Offset Accuracy
Iteration
Analog Static
Value
Amplitude
(V
pp
)
DC Offset
(V)
Ideal DC
Output
(V)
Calibration
Test Limits
(V)
Published
Specification
(V)
1
7FFF
20
0
10
±0.021
±0.102
2
FFFF8000
20
0
–10
±
0.021
±0.102
3
7FFF
17
0
8.5
±0.018
±0.087
4
FFFF8000
17
0
–8.5
±0.018
±0.087
5
7FFF
14
0
7
±0.015
±0.072
6
FFFF8000
14
0
–7
±0.015
±0.072
7
7FFF
10
0
5
±0.011
±0.052
8
FFFF8000
10
0
–5
±0.011
±0.052
9
7FFF
7
0
3.5
±0.008
±0.037
10
FFFF8000
7
0
–3.5
±0.008
±0.037
11
7FFF
5
0
2.5
±0.006
±0.027
12
FFFF8000
5
0
–2.5
±0.006
±0.027
13
7FFF
4
0
2
±0.005
±0.022