NI 1483R User Guide and Specifications
16
ni.com
Storage Environment
Ambient temperature range ...................................–20 °C to 70 °C
Tested in accordance with IEC-60068-2-1
and IEC-60068-2-2
Relative humidity range .........................................5% to 95%, noncondensing,
Tested in accordance with IEC-60068-2-56
Note
Clean the device with a soft, non-metallic brush. Make sure the device is completely dry and
free from contaminants before returning it to service.
Shock and Vibration
Operational shock ..................................................30 g peak, half-sine, 11 ms pulse
Tested in accordance with IEC-60068-2-27.
Test profile developed in accordance with
MIL-PRF-28800F.
Random vibration
Operating .......................................................5 to 500 Hz, 0.3 grms
Nonoperating .................................................5 to 500 Hz, 2.4 grms
Tested in accordance with IEC-60068-2-64.
Nonoperating test profile exceeds the requirements
of MIL-PRF-28800F, Class 3.
Safety
This product meets the requirements of the following standards of safety for electrical equipment for
measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for
measurement, control, and laboratory use:
•
EN 61326 (IEC 61326): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
For the standards applied to assess the EMC of this product, refer to the
section.
Note
For EMC compliance, operate this device with shielded cables and according to the
documentation.
Содержание NI 1483