![National Instruments MXI Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/national-instruments/mxi/mxi_user-manual_709617024.webp)
Appendix A
Specifications
©
National Instruments Corporation
A-3
MXI-Express (x1) for PXI Express Series User Manual
Shock and Vibration
Operational shock .................................. 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating................................................ 5 to 500 Hz, 0.3 g
rms
Nonoperating.......................................... 5 to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC-60068-2-64.
Nonoperating test profile
exceeds the requirements of
MIL-PRF-28800F, Class 3.)
NI PCIe-8361/8362
Maximum altitude .................................. 2,000 m (at 25 °C ambient
temperature)
Pollution Degree .................................... 2
Indoor use only
Operating Environment
Ambient temperature range.................... 0 to 40 °C (Tested in accordance
with IEC-60068-2-1 and
IEC-60068-2-2)
Relative humidity range ......................... 10 to 90%, noncondensing
(Tested in accordance with
IEC-60068-2-56)
Storage Environment
Ambient temperature range.................... –20 to 70 °C (Tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
Relative humidity range ......................... 5 to 95%, noncondensing
(Tested in accordance with
IEC-60068-2-56)