National Instruments IMAQ PXI-1409 Скачать руководство пользователя страница 5

©

 National Instruments Corporation

5

Getting Started with the NI PXI/PCI-1409

Note

Check that the bracket of the device aligns with the hole in the back panel rail of the computer 

chassis.

8.

Secure the device mounting bracket to the back panel rail of the computer.

9.

Replace the computer cover.

10. Connect the BNC cable included with the NI PCI-1409 to the camera. Refer to your camera 

manufacturer documentation for specific instructions about how to connect the cable to your 
camera.

11. Connect the BNC cable to the VIDEO0 connector on the NI PCI-1409 front panel.

12. Plug in and power on the computer.

The NI PCI-1409 is now installed and the camera is connected.

Configuring the NI 1409

After you have installed the NI 1409 and powered on the computer, the computer will recognize the 
device and assign resources to it. Use Measurement & Automation Explorer (MAX), the National 
Instruments configuration utility, to configure the NI 1409 for acquisition. Refer to the 

Measurement & 

Automation Explorer Help for NI-IMAQ

 for additional information about configuring the NI 1409.

Note

Before configuring the device in MAX, ensure that you installed the NI-IMAQ driver 

software.

Specifications

The following specifications apply to the NI 1409 image acquisition device. These specifications are 
typical at 25 °C, unless otherwise stated.

Formats Supported

RS-170/NTSC ........................................................60 Hz (Interlaced mode: 60 fields/s)

CCIR/PAL..............................................................50 Hz (Interlaced mode: 50 fields/s)

VGA.......................................................................60 Hz, 640 

×

 480 resolution

Variable scan ..........................................................Programmable

Video Input

Quantity .................................................................4 monochrome

Input impedance.....................................................75 

Ω ±

1%

VIDEO0 .................................................................RSE (BNC), RSE (VHDCI)

VIDEO<3..0> ........................................................RSE or DIFF (VHDCI)

Frequency response................................................30 MHz (–3 dB) typical

Digital antichrominance filter ................................Programmable (disabled,

3.58 MHz notch filter, or
4.43 MHz notch filter)

Filter characteristics ...............................................Attenuation at notch frequency >30 dB

Input range (black to white)...................................700 mV (calibrated)

50 mV to 1.40 V full scale

Содержание IMAQ PXI-1409

Страница 1: ...Vista NI Vision Acquisition Software 8 0 or later which includes the NI IMAQ driver software Optional software for developing applications NI Vision Builder for Automated Inspection NI Vision Developm...

Страница 2: ...f introducing additional hazards do not install unauthorized parts or modify the device Use the device only with the chassis modules accessories and cables specified in the installation instructions A...

Страница 3: ...ircuit breakers Other examples are wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and stationary motors with permanent connections to fixed installat...

Страница 4: ...or handle to fully inject the device into place 8 Secure the front panel of the NI PXI 1409 to the front panel mounting rails of the PXI or CompactPCI chassis 9 Connect the BNC cable included with the...

Страница 5: ...nt Automation Explorer MAX the National Instruments configuration utility to configure the NI 1409 for acquisition Refer to the Measurement Automation Explorer Help for NI IMAQ for additional informat...

Страница 6: ...eo sources 5 Pixel jitter 2 ns peak PCI Interface PCI initiator master capability Supported PCI target slave capability Supported Data path 32 bits Card voltage 5 V only Card type 32 bit half size car...

Страница 7: ...IL STD 810E Method 514 Test levels exceed those recommended in MIL STD 810E for Category 1 Basic Transportation Figures 514 4 1 through 514 4 3 Approved at altitudes up to 2 000 m Safety This product...

Страница 8: ...assistance from NI Application Engineers A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity Thi...

Отзывы: