Chapter 2
Configuration and Startup Procedures
©
National Instruments Corporation
2-23
lit for an extended period of time. The LED indications are identical for
Non-Slot 0 Message-Based device and Slot 0 Message-Based device
operation.
Slot 0 Message-Based Device Configuration
Follow these steps to configure the GPIB-VXI/C for Slot 0 Message-Based
operation. Refer to Table 2-14.
1.
Enable the VXIbus Slot 0 functions.
2.
Set the model code of the GPIB-VXI/C to be configured for Slot 0
operation using the nonvolatile configuration mode.
3.
Set the logical address to a non-zero value with an appropriate Servant
area size.
If the logical address is set to FFh in nonvolatile configuration (the DIP
switch cannot set the logical address to FFh), the GPIB-VXI/C will
participate in dynamic configuration. Otherwise, the GPIB-VXI/C is a
static configuration device.
Table 2-13.
Front Panel LED Indications for Message-Based Device Operation
Sequence
FAILED
TEST
ON LINE
State
Point of Failure
1
OFF
OFF
OFF
No power
Failed before self-test
2
ON
OFF
OFF
In self-initialization
Failed in self-test
3
ON
ON
OFF
In self-test
—
4
OFF
ON
ON
Performing RM
—
5
OFF
ON
OFF
OFF
OFF
ON
ON
ON
OFF
Online
Failed
In nonvolatile
configuration or
diagnostics mode
—
Failed while online
—
Table 2-14.
Slot 0 Message-Based Device Operation Switch and Jumper Settings
Jumper/Switch
Position
Function
Switches S9
and S15
Refer to Table 2-15.
CLK10 routing options.
Switch S22
ON
VXI BTO enabled.