Test Equipment
The following table lists the equipment required for calibrating your FD-11613 or FD-11614.
If you do not have the recommended instruments, use the minimum requirements to select
substitute equipment.
Table 2. Recommended Test Equipment
Equipment
Recommended Model
Minimum Requirements
Calibrator
Fluke 5522A locked in the 3.3 V
range
A high-precision voltage source with an
uncertainty of ≤70 ppm when sourcing
up to 50 µA.
Mini TC (x8) Omega SMPW-U-M
U type
Test Conditions
The following setup and environmental conditions are required to ensure the FD-11613/11614
meets published specifications:
2
|
ni.com
|
FD-11613/11614 Calibration Procedure