Chapter 5
Calibration
5-14
©
National Instruments Corporation
Calibration DACs
The calibration DACs are used to adjust the errors in the analog signal paths
for errors such as offset and gain. Table 5-7 shows which type of DAC(s)
each board uses for calibration.
394
CALDAC factory constant
AO 1
unipolar
linearity
MB88341
2
1
393
CALDAC factory constant
AO 1
unipolar
gain coarse
MB88341
10
1
392
CALDAC factory constant
AO 1
unipolar
gain fine
MB88341
6
1
391
CALDAC factory constant
AO 1
unipolar
offset
MB88341
14
1
390
Factory calibration temp
—
—
—
1
371
Start of the four user
calibration sections
—
—
—
1
1
Board Codes
(MSB
×
256) + LSB = Board Code
(1
×
256) + 17 = 267
Table 5-7.
Type of CALDAC Used on Board
Board
MB88341
(8-bit)
DAC8800
(8-bit)
DAC8043
(12-bit)
1
AD8522
(12-bit)
2
PCI-MIO-16E-1
✓
—
—
—
PCI-MIO-16E-4
✓
—
—
—
PCI-6071E
✓
—
—
—
PCI-MIO-16XE-50
—
✓
✓
—
PCI-MIO-16XE-10
—
✓
✓
✓
Table 5-6.
PCI-6052E EEPROM Map (Continued)
EEPROM
Address
Data Stored at EEPROM
Address
Data Type
CALDAC
Type
CALDAC
Address
Ser
DacLd
Line