Appendix A
Specifications
DAQMeter DAQCard-4050 User Manual
A-2
©
National Instruments Corporation
Input resistance .................................. >1 G
Ω
(2 V, 200 mV,
20 mV ranges); 1 M
Ω
(250 V,
25 V ranges)
Linearity ............................................ 0.0015%
AC Voltage
Accuracy
Additional AC Errors
Frequency dependent errors
Range
Accuracy
±(% of r
% of range)
Temperature
Coefficient
±(ppm of reading
+ ppm of
range)/°C
Overrange
250.000 V
0.3% + 0.1%
60 ppm + 10 ppm
250.000 V
25.0000 V
0.4% + 0.2%
60 ppm + 80 ppm
27.0000 V
2.00000 V
0.4% + 0.2%
60 ppm + 100 ppm
2.20000 V
200.000 mV
0.3% + 0.1%
60 ppm + 50 ppm
220.000 mV
20.0000 mV
0.3% + 0.1%
60 ppm + 50 ppm
22.0000 mV
Input Frequency
Additional Error
(% of full scale)
20–50 Hz
2%
50–100 Hz
1%
100 Hz–5 kHz
0%
5–10 kHz
1%
10–25 kHz
2%