Troubleshooting Chassis Connectivity
If your cDAQ chassis becomes disconnected from the network, try the following:
•
After moving the chassis to a new network, NI-DAQmx may lose connection to the chassis.
In this case, click
Reconnect
to provide NI-DAQmx with the new hostname or IP address.
•
The cDAQ chassis icon indicates whether it is recognized and present on the network. If a
connected chassis appears as disconnected in the configuration tree in MAX, select
Self-Test
or
Reset Chassis
. If successful, the chassis icon changes to blue/grey.
Figure 1-5.
MAX Icons and States
For additional troubleshooting resources for the cDAQ chassis, refer to the
Finding a Network
DAQ Device in MAX
topic in the
Measurement & Automation Explorer Help for NI-DAQmx
.
Reserving the Chassis in MAX
When the cDAQ chassis is connected to a network, multiple users can access the chassis. To
perform any DAQ functionality on the C Series modules, including reset chassis and self-test,
you must reserve the cDAQ chassis in MAX. Figure 1-5 depicts the chassis state icons in MAX:
an unreserved chassis or chassis reserved by another host appear with an X and reserved chassis
appear as blue/grey. Only one user at a time can reserve the cDAQ chassis.
If the cDAQ chassis was not reserved automatically after it was added (
Add Device
), you can
reserve the cDAQ chassis in MAX by expanding
Devices and Interfaces»Network Devices
,
selecting the chassis, and clicking the
Reserve
Chassis
button. The
Override Reservation
dialog box appears when you attempt to explicitly reserve a chassis. Agreeing to override the
reservation forces the cDAQ chassis to be reserved by the current user.
1
Recognized, but Disconnected from the Network, Unreserved, or Reserved by Another Host
2
Recognized, Present, and Reserved on the Network
1
2