derating. For more information about mounting configurations and temperature
derating, go to
ni.com/info
and enter Info Code
cdaqmounting
.
Pollution Degree (IEC 60664)
2
Maximum altitude
5,000 m
Indoor use only.
Note
Refer to the
NI cDAQ-9132 Specifications
for complete specifications.
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory requirements and limits for
electromagnetic compatibility (EMC) stated in the product specifications. These requirements
and limits provide reasonable protection against harmful interference when the product is
operated in the intended operational electromagnetic environment.
This product is intended for use in industrial locations. However, harmful interference may
occur in some installations, when the product is connected to a peripheral device or test object,
or if the product is used in residential or commercial areas. To minimize interference with
radio and television reception and prevent unacceptable performance degradation, install and
use this product in strict accordance with the instructions in the product documentation.
Furthermore, any changes or modifications to the product not expressly approved by National
Instruments could void your authority to operate it under your local regulatory rules.
Caution
To ensure the specified EMC performance, product installation requires
either special considerations or user-installed add-on devices. Refer to the
NI cDAQ-9132/9133/9134/9135/9136/9137 User Manual
for further information.
Caution
To ensure the specified EMC performance, operate this product only with
shielded cables and accessories. Note that the input DC power cables may be
unshielded.
Caution
To ensure the specified EMC performance, do not connect V2 to a
DC MAINS supply or to any supply requiring a connecting cable longer than 3 m
(10 ft). A DC MAINS supply is a local DC electricity supply network in the
infrastructure of a site or building.
Caution
To ensure the specified EMC performance, the length of any cable
connected to the video and USB host ports must be no longer than 3 m (10 ft). The
length of any cable connected to the RS-232 port must be no longer than 30 m
(100 ft).
Caution
The USB device port is intended for use in device configuration,
application deployment, debug, and maintenance.