NI 9232 Operating Instructions and Specifications
|
© National Instruments
|
29
Input delay
64x decimation ........................... 30/
f
s
+ 3.0
μ
s
128x decimation ......................... 29/
f
s
+ 3.0
μ
s
256x decimation ......................... 28/
f
s
+ 3.0
μ
s
Accuracy
Stability
Gain drift .................................... ±25 ppm/°C
Offset drift (DC coupled) ........... ±320
μ
V/°C
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range
*
(Offset Error
†
)
Calibrated max (-40 °C to 70 °C)
±0.60%
±0.23%
Calibrated typ (23 °C ±5 °C)
±0.10%
±0.023%
Uncalibrated max (-40 °C to 70 °C)
±1.50%
±0.59%
Uncalibrated typ (23 °C ±5 °C)
±0.40%
±0.12%
*
Range = 31.5 V
†
DC coupled