National Instruments Corp.
9
NI 9213E Operating Instructions and Specifications
Figure 5.
Input Circuitry for One Channel of the NI 9213E
The channels share a common ground, COM, that is isolated from
other modules in the system. The NI 9213E common-mode range
is the maximum voltage between any channel and COM. If COM
is not connected, then the common-mode voltage range is the
maximum voltage between any two channels. The NI 9213E
measures the common-mode voltage level of each channel and
returns a warning in the software if the signal is outside the
TC+
TC–
COM
NI 9213E
Input
Impedance
Open
Thermocouple
Detection
Current
Filtered
Differential
Amplifier
and
Multiplexer
Isolated
ADC
39 M
Ω
39 M
Ω