Timing modes
Common-mode voltage range
Channel-to-COM........................ ±1.2 V min
COM-to-earth ground................. ±250 V
Timing Mode
Conversion Time
(Per Channel)
Sample Rate
*
(All Channels
†
)
High-resolution
55 ms
1 S/s
High-speed
740
μ
s
75 S/s
*
If you are using fewer than all channels, the sample rate might be faster.
The maximum sample rate = 1/(
Conversion Time
×
Number of Channels
),
or 100 S/s, whichever is smaller. Sampling faster than the maximum
sample rate may result in the degradation of accuracy.
†
Including the autozero and cold-junction channels.