CALIBRATION PROCEDURE
NI 9211
This document contains the verification and adjustment procedures for the National Instruments
9211. For more information about calibration solutions, visit
ni.com/calibration
.
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 2
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 3
EEPROM Update ..................................................................................................................... 6
Re-Verification ......................................................................................................................... 6
Accuracy Under Calibration Conditions .................................................................................. 6
Where to Go for Support .......................................................................................................... 7
Software
Calibrating the NI 9211
requires the installation of NI-DAQmx 9.2 or later
on the calibration
system. You can download NI-DAQmx from
ni.com/downloads
. NI-DAQmx supports
LabVIEW, LabWindows
™
/CVI
™
, C++, and .NET. When you install NI-DAQmx, you only need
to install support for the application software that you intend to use.