10
|
ni.com
|
NI 6738/6739 Calibration Procedure
2.
Measure the internal device temperature and perform self-calibration if necessary. Refer to
the
Checking Device Temperature Changes
section for more information.
3.
Call
DAQmxCreateTask
with the following parameters:
•
taskName
:
MyAOVoltageTask
•
taskHandle
:
&taskHandle
4.
Call
DAQmxCreateAOVoltageChan
with the following parameters.
•
taskHandle
:
taskHandle
•
physicalChannel
:
Dev_name/ao0
•
nameToAssignToChannel
:
AOVoltageChannel
•
minVal
:
-10.0
•
maxVal
:
10.0
•
units
:
DAQmx_Val_Volts
•
customScaleName
:
NULL
5.
Call
DAQmxStartTask
with the following parameter:
•
taskHandle
:
taskHandle
6.
Call
DAQmxWriteAnalogF64
with the following parameters, using a
Test Point
from
Table 3,
NI 6738/6739 Analog Output Values
•
taskHandle
:
taskHandle
•
numSampsPerChan
:
1
•
autoStart
:
1
•
timeout
:
10.0
•
dataLayout
:
DAQmx_Val_GroupByChannel
•
writeArray
:
&data
AO 57
Connector 1, AO 57 (pin 63)
Connector 1, AO GND 56/57 (pin 30)
AO 58
Connector 1, AO 58 (pin 64)
Connector 1, AO GND 58/59 (pin 65)
AO 59
Connector 1, AO 59 (pin 31)
Connector 1, AO GND 58/59 (pin 65)
AO 60
Connector 1, AO 60 (pin 32)
Connector 1, AO GND 60/61 (pin 33)
AO 61
Connector 1, AO 61 (pin 66)
Connector 1, AO GND 60/61 (pin 33)
AO 62
Connector 1, AO 62 (pin 67)
Connector 1, AO GND 62/63 (pin 68)
AO 63
Connector 1, AO 63 (pin 34)
Connector 1, AO GND 62/63 (pin 68)
*
Connector 1/AO <32..63> are on the PXIe-6739 only.
Table 2.
Analog Output Connections (Continued)
Analog
Output
*
DMM
Positive Input
Negative Input