NI USB-6008/6009 User Guide and Specifications

16

ni.com

Analog Output

The NI USB-6008/6009 has two independent AO channels that can 
generate outputs from 0–5 V. All updates of AO lines are software-timed.

Analog Output Circuitry

Figure 11 illustrates the analog output circuitry for the NI USB-6008/6009. 

Figure 11.  

Analog Output Circuitry

DACs

Digital-to-analog converts (DACs) convert digital codes to analog 
voltages.

Connecting Analog Output Loads

To connect loads to the NI USB-6008/6009, connect the positive lead of the 
load to the AO terminal, and connect the ground of the load to a GND 
terminal. 

Figure 12.  

Connecting a Load

REF(+) REF(–)

12-Bit

DAC

Output

Buffer

50

 Ω

AO

+5 V

GND

AI

Volt

a

ge

S

o

u

rce

NI U

S

B-600

8

/6009

GND

Содержание 6008

Страница 1: ...nput AI channels two analog output AO channels 12 digital input output DIO channels and a 32 bit counter with a full speed USB interface Note This manual revision updates naming conventions to reflect the conventions used in NI DAQmx Table 1 notes the correlation between the old and updated names Table 1 Digital Output Driver Type Naming Conventions Hardware Functionality NI DAQmx Terminology Open...

Страница 2: ... 6009 AI Resolution 12 bits differential 11 bits single ended 14 bits differential 13 bits single ended Maximum AI Sample Rate Single Channel 10 kS s 48 kS s Maximum AI Sample Rate Multiple Channels Aggregate 10 kS s 48 kS s DIO Configuration Open collector Open collector or active drive System dependent 1 USB Cable Strain Relief NI USB 6009 8 Inputs 14 bit Multifunction I O 32 Digital 17 1 Analog...

Страница 3: ... If you need to clean the device use a dry cloth Make sure that the device is completely dry and free from contaminants before returning it to service Operate the device only at or below Pollution Degree 2 Pollution is foreign matter in a solid liquid or gaseous state that can reduce dielectric strength or surface resistivity The following is a description of pollution degrees Pollution Degree 1 m...

Страница 4: ...ard wall outlet for example 115 V for U S or 230 V for Europe Examples of Measurement Category II are measurements performed on household appliances portable tools and similar E Series devices Measurement Category III is for measurements performed in the building installation at the distribution level This category refers to measurements on hard wired equipment such as equipment in fixed installat...

Страница 5: ... Guide that shipped with your device and is also accessible from Start All Programs National Instruments NI DAQ for more information Note For information about non Windows operating system support refer to ni com info and enter rddqld LabVIEW SignalExpress for DAQ The NI DAQmx CD includes LabVIEW SignalExpress for DAQ which is an interactive measurement software tool for quickly acquiring analyzin...

Страница 6: ... functional components of the NI USB 6008 6009 Figure 3 Device Block Diagram P1 0 3 P0 0 7 Digital I O Terminal Block AI 0 7 AO 0 AO 1 Analog I O Terminal Block 2 5 V CAL 5 V 200 mA 8 Channel 12 14b ADC 12b DAC 12b DAC USB Microcontroller SPI External Power Supply Vbus Full Speed USB Interface USB PFI 0 ...

Страница 7: ...can apply the signal labels to the screw terminal blocks for easy signal identification Figure 4 NI USB 6008 6009 Signal Labels 3 Refer to Table 3 and Figures 4 and 5 for signal label orientation and affix the provided signal labels to the screw terminal blocks Until the signal labels are applied you can insert the screw terminal blocks into either of the combicon jacks 1 Terminal Number Labels Us...

Страница 8: ...hem into the matching combicon jack as indicated by the overlay label on the NI USB 6008 6009 device 4 Connect the wiring to the appropriate screw terminals 1 Overlay Label with Pin Orientation Guides 2 Combicon Jack 3 Signal Labels 4 USB Cable 9 0 0 6 B S U I N O I n o i t c n u f i t l u M t i b 4 1 s t u p n I 8 2 3 l a t i g i D 7 1 1 g o l a n A 6 1 3 2 4 1 2 3 ...

Страница 9: ...de 16 connections that use 16 AWG to 28 AWG wire Table 3 lists the analog terminal assignments and Table 4 lists the digital terminal assignments Table 3 Analog Terminal Assignments Module Terminal Signal Single Ended Mode Signal Differential Mode 1 GND GND 2 AI 0 AI 0 3 AI 4 AI 0 4 GND GND 5 AI 1 AI 1 6 AI 5 AI 1 7 GND GND 8 AI 2 AI 2 9 AI 6 AI 2 10 GND GND 11 AI 3 AI 3 12 AI 7 AI 3 13 GND GND 14...

Страница 10: ...s 10 ni com Table 4 Digital Terminal Assignments Module Terminal Signal 17 P0 0 18 P0 1 19 P0 2 20 P0 3 21 P0 4 22 P0 5 23 P0 6 24 P0 7 25 P1 0 26 P1 1 27 P1 2 28 P1 3 29 PFI 0 30 2 5 V 31 5 V 32 GND 32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 ...

Страница 11: ...ltage channel For differential measurements AI 0 and AI 4 are the positive and negative inputs of differential analog input channel 0 The following signal pairs also form differential input channels AI 1 AI 5 AI 2 AI 6 and AI 3 AI 7 AO 0 GND Output Analog Channel 0 Output Supplies the voltage output of AO channel 0 AO 1 GND Output Analog Channel 1 Output Supplies the voltage output of AO channel 1...

Страница 12: ... has NI DAQmx installed on it If your device is not blinking make sure your computer has the latest version of NI DAQmx installed on it and the computer is not in standby mode Analog Input You can connect analog input signals to the NI USB 6008 6009 through the I O connector Refer to Table 5 for more information about connecting analog input signals Analog Input Circuitry Figure 6 illustrates the ...

Страница 13: ... AI FIFO The NI USB 6008 6009 can perform both single and multiple A D conversions of a fixed or infinite number of samples A first in first out FIFO buffer holds data during AI acquisitions to ensure that no data is lost Analog Input Modes You can configure the AI channels on the NI USB 6008 6009 to take single ended or differential measurements Refer to Table 5 for more information about I O con...

Страница 14: ...o GND For example if AI 1 is 10 V and AI 5 is 10 V then the measurement returned from the device is 20 V Figure 8 Example of a Differential 20 V Measurement Connecting a signal greater than 10 V on either pin results in a clipped output Figure 9 Exceeding 10 V on AI Returns Clipped Output 5 10 15 20 20 15 10 5 0 Amplitude V AI 1 AI 5 Result 5 10 15 20 20 15 10 5 0 Amplitude V AI 1 AI 5 Result ...

Страница 15: ...se the terminal to float to approximately 1 4 V when the analog input terminal is configured as RSE This behavior is normal and does not affect the measurement when a signal is connected Figure 10 Connecting a Reference Single Ended Voltage Signal Digital Trigger When an AI task is defined you can configure PFI 0 as a digital trigger input When the digital trigger is enabled the AI task waits for ...

Страница 16: ...nalog output circuitry for the NI USB 6008 6009 Figure 11 Analog Output Circuitry DACs Digital to analog converts DACs convert digital codes to analog voltages Connecting Analog Output Loads To connect loads to the NI USB 6008 6009 connect the positive lead of the load to the AO terminal and connect the ground of the load to a GND terminal Figure 12 Connecting a Load REF REF 12 Bit DAC Output Buff...

Страница 17: ...litches Digital I O The NI USB 6008 6009 has 12 digital lines P0 0 7 and P1 0 3 which comprise the DIO port GND is the ground reference signal for the DIO port You can individually program all lines as inputs or outputs Digital I O Circuitry Figure 13 shows P0 0 7 connected to example signals configured as digital inputs and digital outputs You can configure P1 0 3 similarly Figure 13 Example of C...

Страница 18: ...ctive drive using the DAQmx API allowing 3 3 V operation with a source sink current limit of 8 5 mA Refer to the NI DAQmx Help for more information about how to set the DIO configuration Figure 14 Example of Connecting External User Provided Resistor Complete the following steps to determine the value of the user provided pull up resistor 1 Place an ammeter in series with the load 2 Place a variab...

Страница 19: ...igh current drive If you configure a DIO line as an input do not drive the line with voltages outside of its normal operating range The DIO lines have a smaller operating range than the AI signals Treat the DAQ device as you would treat any static sensitive device Always properly ground yourself and the equipment when handling the DAQ device or connecting to it Power On States At system startup an...

Страница 20: ...test 5 V Power Source The NI USB 6008 6009 supplies a 5 V 200 mA output This source can be used to power external components Note While the device is in USB suspend the output is disabled Specifications The following specifications are typical at 25 C unless otherwise noted Analog Input Converter type Successive approximation Analog inputs 8 single ended 4 differential software selectable Input re...

Страница 21: ...s Differential 20 V range 5 mVrms 1 V range 0 5 mVrms Absolute accuracy at full scale single ended Absolute accuracy at full scale differential3 1 20 V means that AI AI 20V However AI and AI must both be within 10V of GND Refer to the Connecting Differential Voltage Signals for more information 2 System noise measured at maximum sample rate Range Typical at 25 C mV Maximum over Temperature mV 10 1...

Страница 22: ...cuit current 50 mA Absolute accuracy no load 7 mV typical 36 4 mV maximum at full scale Digital I O Digital I O P0 0 7 8 lines P1 0 3 4 lines Direction control Each channel individually programmable as input or output Output driver type NI USB 6008 Open collector open drain NI USB 6009 Each channel individually programmable as active drive push pull or open collector open drain 2 2 21 42 5 1 25 1 ...

Страница 23: ...erature drift 50 ppm C max Counter Number of counters 1 Resolution 32 bits Counter measurements Edge counting falling edge Counter direction Count up Pull up resistor 4 7 kΩ to 5 V Maximum input frequency 5 MHz Minimum high pulse width 100 ns Minimum low pulse width 100 ns Level Min Max Units Input low voltage Input high voltage Input leakage current 0 3 2 0 0 8 5 8 50 V V μA Output low voltage I ...

Страница 24: ...l 500 μA max Physical Characteristics If you need to clean the module wipe it with a dry towel Dimensions Without connectors 6 35 cm 8 51 cm 2 31 cm 2 50 in 3 35 in 0 91 in With connectors 8 18 cm 8 51 cm 2 31 cm 3 22 in 3 35 in 0 91 in I O connectors USB series B receptacle 2 16 position terminal block plug headers Weight With connectors 84 g 3 oz Without connectors 54 g 1 9 oz Screw terminal wir...

Страница 25: ...s not directly connected to the electrical distribution system referred to as MAINS voltage MAINS is a hazardous live electrical supply system that powers equipment This category is for measurements of voltages from specially protected secondary circuits Such voltage measurements include signal levels special equipment limited energy parts of equipment circuits powered by regulated low voltage sou...

Страница 26: ...tion after RF exposure is removed Note For EMC compliance operate this device according to product documentation CE Compliance This product meets the essential requirements of applicable European Directives as amended for CE marking as follows 73 23 EEC Low Voltage Directive safety 89 336 EEC Electromagnetic Compatibility Directive EMC Note Refer to the Declaration of Conformity DoC for this produ...

Страница 27: ...uct supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com sup...

Отзывы: