This CLIP also provides a User Command Interface to configure the clock selection (Internal
Sample Clock Free Running, Internal Sample Clock locked to an external Reference Clock
through the CLK/REF IN connector, 60 to 100 MHz External Sample Clock through the
CLK/REF IN connector, or Internal Sample Clock locked to an external Reference Clock
through Sync Clock), read the CLIP version, reinitialize the adapter module, and send raw SPI
and I
2
C accesses to the adapter module for low-level configuration.
Related Information
NI 5783 400 MSPS AO CLIP
This CLIP provides access to four 16-bit 100 MS/s analog input channels, two 16-bit
400 MS/s analog output channels, eight bidirectional DIO channels, four bidirectional PFI
channels, and the front panel trigger.
The analog output channels can be written four samples at a time in the Data Clock domain
with 2x interpolation, or two samples at a time in the Data Clock 2x domain with 2x
interpolation. In addition, the two analog output channels are provided through channels 1 and
2 of the front panel.
This CLIP also provides a User Command Interface to configure the clock selection (Internal
Sample Clock Free Running, Internal Sample Clock locked to an external Reference Clock
through the CLK/REF IN connector, 60 to 100 MHz External Sample Clock through the
CLK/REF IN connector, or Internal Sample Clock locked to an external Reference Clock
through Sync Clock), and send raw SPI and I
2
C accesses to the adapter module for low-level
configuration.
Related Information
Clocking
The following clock modes are available on the NI 5783.
•
Internal Sample Clock Free Running
•
Internal Sample Clock locked to an external Reference Clock through the CLK/REF IN
front panel connector
NI 5783 Getting Started Guide
|
© National Instruments
|
11