B. LUNG BLOCKS
1.
Two different lung blocks have been pro-
vided: a normal block simulating bilateral
expansion of the lungs, and a second block
simulating a collapsed right lung. The two
blocks can be identified by the molded
diagram on the back.
2.
To install either lung block, ensure the
airway is installed (
See “A. Installing the
Replaceable Airway” for more information)
.
3.
Apply a small amount of lubricant to the
short tubing section of the airway that projects into the chest cavity.
4.
Insert the lubricated tubing section into the hole at the top of the lung block.
(See figure 3.)
5.
Fit the lung block into the chest cavity.
C. PATENT UMBILICUS
The umbilical stump has one functional vessel and functions like a plug that fits into the small reservoir
within the abdomen.
D. OSTOMY
The nonfunctional stoma can be positioned externally, or pulled out and reinserted with the skin side up.
GENERAL INSTRUCTIONS FOR USE
AIRWAY AND VENTILATION
The baby will accept a functioning 2.5 mm I.D. endotracheal tube (not included) with #0 Miller Blade.
The lubricated tube must be placed into the throat to produce a visible chest rise. Do not administer
fluids through the mouth. BVM can also be used to produce a chest rise.
GI
The left nostril will accept a 5 Fr nasogastric feeding tube (not provided). Fluids passed through the NG
tube will enter an embedded tube located under the chest block.
Note:
Fluid administration is not recommended. Any fluid administered must be suctioned out. The
embedded tube is approximately 1-2 cc. Use water only, not formula. Always drain the system completely
when training is complete.
6
3
Figure 3
UMBILICUS
The reservoir can be used to either draw or receive fluids. Maximum capacity is approximately 5 cc.
Do not overfill. Always use plain water, with or without the blood powder provided.
(
Caution:
Blood mixture will stain skin and fabric.)
(See figures 4 & 5.
) Always flush and drain the system
thoroughly and store with the plug hole open so the reservoir can dry.
Figure 4
Figure 5
NG Access with
5 cc FR NG Tube
ET Access with
#0 Miller and
2.5 mm ET Tube
Lung Cavity
Nonfunctional
Vein Sites
Umbilical
Reservoir
Nonfunctional
Vein
Nonfunctional
Scalp Vein Site
Chest Tube
Stoma
HC18000028INM32.indd 5-6
10/9/18 3:09 PM