background image

14

7

4. 

Attach fitting end of tubing from bag A to the shoulder tubing that has the pinch clamp. Attach the 

fitting end of bag B to the remaining shoulder tube.

5. 

With bag A hanging and bag B laying on the surface, open clamps on both bags and the arm tube 

clamp. Allow fluid to flow until air bubbles are gone. The venous system is now pressurized. Close 

the clamp on bag B.

6. 

Insert IV needle (or butterfly) in vein. “Flashback” will indicate proper insertion.

7. 

Close the clamp on bag A and the clamp on the arm tube. Disconnect bag A from the shoulder 

tubing. You may now use bag A as the infusion supply.

8. 

Attach the needle or butterfly to bag A using the latex adapter. 

(See figure 3.)

9. 

To start the IV flow, open the clamps on both bags A and B.

Proof of proper procedure will be evidenced by the flow of fluid from bag A. Control flow rate with the 

clamp on bag A. This fluid can be reused. A user-supplied IV bag may also be used for the infusion fluid. 

This will enable bags A and B to remain attached to the arm. If a more realistic experience is desired with 

“blood flashback” instead of water when inserting butterfly into lumen of vein, use procedure C.

C.  Recommended Procedure for Simultaneous IV Infusions and Drawing “Blood”

Use two IV Bag Kits. Hook up and install with IV bag A and IV bag B.

 (See figure 2.)

1. 

Begin with synthetic blood in IV bag A. Open the clamps on both A and B to pressurize the system. 

“Flush” system by allowing “blood” to flow into container B until bubbles in tubing disappear, then 

regulate blood flow from bag A (using the clamp). System is now full of “blood” and pressurized. 

“Blood” can now be drawn anywhere along the pathway of the vein.

2. 

Intravenous infusion — Insert the butterfly needle into the lumen of the vein: A flashback of “blood” 

is proof the needle was inserted correctly. Close the clamp on the bag A tubing and disconnect it at 

the shoulder. Use the extra pinch clamp supplied to clamp off the loose shoulder tubing. Connect 

the IV tubing from bag A to the butterfly needle using the special connector supplied. Open the 

clamp on the bag A tubing and adjust it as desired. To keep the process going, when bag B fills, 

simply switch the positions of bags A and B and their IV lines. 

Note: 

Always regulate the flow of “blood” from the bag on the stand, and be sure the clamp on the 

other bag is open.

Causes for Failure in Function

A. 

Forgetting to open a clamp.

B. 

Kinks in tubing of IV sets.

C. 

Is the supply hung at the appropriate height according to the instruction manual? Hanging the 

supply bags “slightly” higher for the bags that are not producing enough pressure can create just 

enough gravity on the fluid to facilitate flow.

D. 

Tubing pinched shut by constant pressure of clamps. Lumen remains pinched occasionally, even if 

clamp is loosened. Slide clamp to new position and manipulate tubing with fingers at pinched site 

to restore lumen. In heavy use, slide clamp to new position on tubing from time to time to prevent 

the “permanent pinch” caused by constant clamp pressure. Replace IV kit.

ABOUT THE SIMULATOR

The 

Life

/

form

®

 Child Defibrillation Chest Skin has been designed for the Child 

CR

i

S

is

 manikin to 

enable you, the customer, to safely practice defibrillation. The Child Defibrillation Chest Skin has been 

designed to absorb a maximum of 360 joules of energy*. Although capable of absorbing 360 joules, we 

recommend that the smallest energy level possible be used while training with the skin.
The Child Defibrillation Chest Skin will enable you to practice defibrillation using manual, semiautomatic, 

and automatic external defibrillators (AEDs). When using any one of these types of defibrillators in train-

ing, always follow the recommended operating procedures for that particular defibrillator. 

LIST OF COMPONENTS

• 

Defibrillation Chest Skin with Load Box and Four Lead Snap Cable Connector. 

(See figure 1.)

*Note: 

360 joules is the maximum energy level that Nasco recommends administering to the defibril-

lation chest skin. Energy levels in excess of 360 joules may cause irreparable damage to the chest skin, 

circuitry, and patient simulator being used — thus voiding Nasco’s warranty and endangering your 

equipment. Nasco assumes no liability for damage or injury that may be caused by the use and/or misuse 

of this equipment. All normal safety precautions for defibrillation training should be followed, and energy 

levels should be minimized. Nasco did not design nor intend this defibrillation chest skin to be used as 

anything other than a training apparatus for defibrillation.

CONNECTING YOUR PATIENT SIMULATOR

Connect the four lead snap cable to your patient simulator.
Nasco has designed the Child 

CR

i

S

is

 System to be compatible with a variety of patient simulators. 

This is possible via the standard four lead snap cable.

 (See figure 2.)

 If your patient simulator has only two 

output posts, the red and black leads must be connected to the patient simulator.
If you want to connect the manikin to the simulator that came with your defibrillator, it is necessary to 

purchase the corresponding adapters from Nasco separately, if the patient simulator doesn’t have the 

standard snap connectors. (Please see accessories at the end of this section).

THE DEFIBRILLATION CHEST SKIN

Figure 1

Figure 3

Figure 2

HC190001INM14.indd   13-14

2/20/19   11:23 AM

Содержание CRiSis LF03616U

Страница 1: ...HEALTHCARE ChildCRiSis Manikin Simulator LF03616U INSTRUCTION MANUAL WARNING Products may contain dry natural rubber 5 Year Warranty erials ...

Страница 2: ...2 TABLE OF CONTENTS Introduction 3 Airway Management Trainer 4 IV Arm 5 Blood Pressure Arm 9 Defibrillation Chest Skin 14 Intraosseous Infusion Femoral Access Leg 16 ...

Страница 3: ...al access ASSEMBLY OF YOUR CHILD CRiSis MANIKIN To facilitate packing your manikin comes with the arms and legs unattached To connect the arms to the torso place each in a position such that the hand is above the head See figure 1 Align the arm fitting with the torso socket and plug the arm into place Finish by rotating the arm forward so it lies along the side of the body To connect the legs to t...

Страница 4: ...te both the oral cavity and the endotracheal tube with the pump spray lubricant provided See figure 1 Note Nasco recommends the use of the provided lubricant or a similar vegetable based lubricant for the Child Trainer The use of a silicone based or similar lubricant may seriously damage the simulator thus voiding Nasco s warranty on the trainer CLEANING THE FACE Normal soil can be removed from th...

Страница 5: ...eplacement Kits are available through Nasco see list of supplies LIST OF COMPONENTS 3 cc Syringe with Needle 12 cc Syringe with Needle 2 IV Bags Needle Butterfly 1 Pinch Clamp Small Towel Arm INTERNAL STRUCTURE Internally the vascular structure rubber tubing begins at the shoulder and continues under the arm crosses the antecubital fossa forearm makes a loop in the back of the hand and then return...

Страница 6: ...fter filling the venous system according to instructions the arm is now ready to practice drawing blood Blood can be drawn anywhere along the pathway of the vein See figure 1 Distilled water should be used to prepare the sites Synthetic blood will actually be aspirated once the vein is properly punctured 8 Small diameter needles 22 gauge or smaller should be used B Preparing the Arm for Intravenou...

Страница 7: ...flow into container B until bubbles in tubing disappear then regulate blood flow from bag A using the clamp System is now full of blood and pressurized Blood can now be drawn anywhere along the pathway of the vein 2 Intravenous infusion Insert the butterfly needle into the lumen of the vein A flashback of blood is proof the needle was inserted correctly Close the clamp on the bag A tubing and disc...

Страница 8: ...occurs Stubborn stains may be removed with Nasco Cleaner LF09919U simply by dispensing it on the area and wiping with a soft cloth or paper towel CAUTIONS 1 This synthetic blood is specially formulated to be compatible with the self sealing veins and plastics used in manufacturing the arm 2 DO NOT use dull or burred needles as these will cause leaks in the system Burred needles will cause permanen...

Страница 9: ...control box has a battery saving feature which will turn the unit off after about 8 10 minutes if no keys are used within that period of time Figure1 ective set been that the g s AMP e the ce he eing le he perly oma ssible rm ssure mHg at dow s 2 ed at 13 BRA Figure2 On Off Switch The next step is to connect the simulated arm and speaker assembly along with the sphygmomanometer cuff and gauge asse...

Страница 10: ...utomatically be set to 0 Figure5 Figure6 Figure7 Figure8 PALPABLE PULSE FEATURE The Nasco Life form Child Blood Pressure Simulator also incorporates a palpable pulse at the radial location See figure 9 Palpations can be felt upon start up of the unit or after blood pressure settings have been made Press the Menu key repeatedly until Set PALPATION menu appears Pulse ON is defaulted and enables the ...

Страница 11: ...henomenon Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the auscultatory gap is enabled or disabled Also the main display will show at the bottom right of the display either AGap Y for on or AGap N for off The arrow up and down keys also control the volume of the sounds that are present in the arm From the main menu press th...

Страница 12: ...tion of the measurement sequence LOW BATTERY INDICATOR When the battery supply diminishes to a level near the point that the unit will no longer function properly the low batt segment of the systolic pressure display will activate when the pressure in the sphygmoma nometer cuff reaches above 20 mmHg At this point the batteries should be replaced as soon as possible in order to insure proper operat...

Страница 13: ...e form CHILD BLOOD PRESSURE SIMULATOR To adapt your sphygmomanometer for use with the simulator it is first necessary to obtain the luer fitting and the T fitting included with the simulator which is also available through your Nasco catalog sales office Using a scissors carefully cut the pressure line of the sphygmomanometer about 2 from the gauge See figure 14 Take the T fitting and insert it be...

Страница 14: ...st skin Energy levels in excess of 360 joules may cause irreparable damage to the chest skin circuitry and patient simulator being used thus voiding Nasco s warranty and endangering your equipment Nasco assumes no liability for damage or injury that may be caused by the use and or misuse of this equipment All normal safety precautions for defibrillation training should be followed and energy level...

Страница 15: ... connections on the patient simulator one or more may be disconnected 2 Check to make sure your patient simulator is plugged in and working properly Problem ECG wave is inverted Solution Recheck the position of the red and black lead snaps on the patient simulator Note If the defibrillation chest skin is not functioning or wiring comes undone please contact us to repair or replace the unit Failure...

Страница 16: ...1 60 cc Syringe K Y Jelly Wax 1 2 oz Liquid Lubricant 2 Leg Skins 2 Mixing Bottles w Blood Mix IV Bag THE INTRAOSSEOUS INFUSION FEMORAL ACCESS LEG GENERAL INSTRUCTIONS FOR INTRAOSSEOUS INFUSION USE 1 Place the simulator on a flat surface such as a tabletop Lay towels under the knees for support and to absorb any overflow 2 To prepare the blood mixture combine the red coloring 1 full tube of K Y Je...

Страница 17: ...ed Extreme caution should be taken to avoid pushing the needle completely through the simulator injuring the person performing the proce dure Nasco cannot be responsible for injuries resulting from improper use of the simulator ure1 4 Connect the end of the tubing to the bone piece Apply liquid lubricant to the entire bone and slide into posi tion in the leg See figure 3 IMPORTANT 5 Make your firs...

Страница 18: ...e access site tubing See figure 6 4 Open the clamp on the IV bag 5 To charge the vein make your first needle inser tion in the vein and pull back on the syringe to remove air from the vein Repeat until blood is withdrawn The vein is now charged and subse quent insertions will produce an immediate flow of blood CLEAN UP PROCEDURES 1 Close the clamp on the IV bag pull the fitting from the femoral ac...

Страница 19: ...iSis Manikin with Arrhythmia Tutor LF03709U Infant CRiSis Manikin LF03953U CRiSis Manikin Complete LF03955U Deluxe CRiSis Manikin LF03956U Deluxe Plus CRiSis Manikin LF03965U Adult CRiSis Auscultation Manikin LF03966U Adult CRiSis Auscultation Manikin with ECG Simulator LF03709U LF03965U Actual product may vary slightly from photo Nasco reserves the right to change product color materials supplies...

Отзывы: