DISASSEMBLY
To remove the contaminate lung system and replace it:
1.
Remove mouth/nose piece from face.
2.
Remove chest skin from torso by detaching
strap snaps.
3.
Remove or displace the chest plate.
4.
Remove and discard the disposable lung.
5.
Remove head.
a.
Pull cotter pin from rod.
b.
Pull out the rod.
c.
Lift the head.
d.
Detach the airway tube from the ball valve
by twisting the airway valve while pulling it
away from the ball valve base.
6.
Pull airway tube from chest cavity.
7.
Remove the ball valve assemble from the head
of necessary.
a.
Unscrew the white nylon nut (will require
wrench or pliers).
ELECTRONIC MODEL ONLY
It is necessary to access the interior of the chest cavity in this model when you are changing the airway
tubing. Separate the pelvis from the torso.
1.
Remove the hair cotter pin at the waist.
2.
Slide the rod out.
3.
The torso and pelvis will now separate with a gentle tug of the pelvis away from the torso.
To attach the torso to the pelvis, take these steps:
1.
Align body parts.
2.
Insert shaft from the left side.
3.
Install washer and hairpin.
CLEANING & DISINFECTING
The only components of Simulaids’ Full Body CPR Manikin that require scrupulous disinfection are the
airway cutoff ball valve and the mouth/nose pieces. The airway tubing and valve are designed to be
disposed of at the end of the training class. however, they may be reused after carrying out the following
procedure:
1.
Clean in mild soap and water.
2.
Rinse thoroughly in clear water.
3.
Disinfect as per CDC protocols.
4.
Rinse thoroughly and dry.
REASSEMBLY
1.
Secure the ball valve into head, making sure
that the longitude of the ball valve pipe is
oriented straight and parallel with the center
line of the head.
2.
Replace nylon nut if you removed it. Make sure that the alignment is correct and tighten the nut
securely with a wrench or pliers to hold the ball valve securely in position.
3.
Insert the one way valve located at the top of the airway into the opened port of the ball valve inside
the head.
4.
Feed tubing down through hole in neck.
5.
Re-attach head with clevis pin and hair cotter pin.
6.
Attach lung to airway tube.
7.
Reposition chest plate and secure with strap snap.
8.
Reposition overlay and attach snaps of the waist strap and the shoulder straps by crossing them to
the opposite side of the body.
9.
Attach squeeze bulb assembly if carotid pulses are desired. Twist the tubing when inserting it into
the neck connection. This will assist it in staying in place.
ELECTRONIC MODEL REASSEMBLY
1.
Reach into chest cavity and feed airway tubing through the hole in chest cavity.
2.
Attach the lung to airway tube.
3.
Fit upper and lower torso together.
a.
Feed the rod through the waist tube.
b.
Position washer over the end of the rod.
c.
Attach hair cotter pin.
4.
Put chest plate and chest skin in place.
CAROTID PULSE
To activate the pulse, plug the connector from the squeeze
bulb assembly into the portal behind the left ear. Rhythmically
squeeze to simulate the pulse.