18 Annex A
208
SRM-3006
Narda
18.1
Pendulum method
Measurement setup
The antenna is mounted directly on the instrument or the antenna is
connected to the Basic Unit via a short cable.
Information about connecting an antenna is found under
antenna directly on the Basic Unit on page 30
antenna to the Basic Unit using a Narda cable on page 31
.
Procedure
During the measurement, the user traverses the space being measured.
The entire volume of space to be measured is sampled slowly by means of
the hand-held antenna, which is moved up and down on a wave-shaped
path (i.e. swung, like a pendulum) through the room, changing the direction
of the main lobe and the direction of polarization at the same time.
During this process, regardless of the selected operating mode, the SRM
should be set to result type
MAX
(
Maximum hold
).
Experience has shown that a few minutes are usually sufficient time for
determining the maximum field strength occurring within a specific
measurement volume such as a room in a house or an office.
Advantages of this method
• Immediate determination of the maximum emission value for the room
being measured
• Takes only a short time
• Minimal disturbance to the use of the rooms affected
• No special preparation for the measurement needed (e.g. moving
furniture, etc.)
Useful application
• Rapid determination of the maximum field strength in a room
• Searching for maximum values in areas where tolerance to disruption is
low (e.g. offices)
• Searching for maximum values at locations where a high degree of
mobility is needed (e.g. on platforms, roofs)
• Sufficient experience is necessary
Содержание SRM-3006
Страница 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Страница 12: ...Contents X SRM 3006 Narda ...
Страница 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Страница 24: ...1 Introduction 12 SRM 3006 Narda ...
Страница 49: ...3 Connecting Up and Starting to Use the Device Narda SRM 3006 37 Figure 6 Cable and N connector in guideway ...
Страница 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Страница 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Страница 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Страница 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Страница 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Страница 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Страница 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Страница 186: ...15 PC Software 174 SRM 3006 Narda ...
Страница 192: ...16 Remote Control 180 SRM 3006 Narda ...
Страница 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Страница 218: ...17 Specifications 206 SRM 3006 Narda ...
Страница 224: ...18 Annex A 212 SRM 3006 Narda ...
Страница 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Страница 231: ...20 Glossary Narda SRM 3006 219 20 Glossary The meanings of terms and abbreviations are explained in this chapter ...
Страница 238: ...Index 226 SRM 3006 Narda ...
Страница 239: ......