10 Scope Mode
Narda
SRM-3006
129
Procedure for example 3:
1.
Switch to
Spectrum Analysis
mode, select a suitable frequency using
the marker function, and then switch to
Scope
mode (parameters are
retained)
ª
Frequency hopping can make this selection more difficult.
2.
Examples of different settings, RBW = 20 MHz:
Spectrum Analysis mode
Frequency hopping
• Sweep Time = 1 ms
• Time resolution = 1/RBW =
1/20 MHz = 50 ns
–> corrected to 400 ns
• Sweep Time = 5 ms
• Time resolution = 1400 ns
Содержание SRM-3006
Страница 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Страница 12: ...Contents X SRM 3006 Narda ...
Страница 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Страница 24: ...1 Introduction 12 SRM 3006 Narda ...
Страница 49: ...3 Connecting Up and Starting to Use the Device Narda SRM 3006 37 Figure 6 Cable and N connector in guideway ...
Страница 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Страница 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Страница 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Страница 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Страница 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Страница 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Страница 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Страница 186: ...15 PC Software 174 SRM 3006 Narda ...
Страница 192: ...16 Remote Control 180 SRM 3006 Narda ...
Страница 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Страница 218: ...17 Specifications 206 SRM 3006 Narda ...
Страница 224: ...18 Annex A 212 SRM 3006 Narda ...
Страница 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Страница 231: ...20 Glossary Narda SRM 3006 219 20 Glossary The meanings of terms and abbreviations are explained in this chapter ...
Страница 238: ...Index 226 SRM 3006 Narda ...
Страница 239: ......