10.1.3 Display in % of standard
Set the reference frequency:
1.
Select
Reference Frequency
.
↳
A list opens with the calibration frequencies stored in the probe. The last selected
value is automatically highlighted.
2.
If necessary, select another value from the list.
In addition to the values from the list, any intermediate value can be set with a resolution
of 1 kHz.
Set any value:
1.
After selecting
Reference Frequency
, press the
SK1
softkey.
2.
Set the digit and value with the cross key.
3.
Press
SK4
to transfer the new value to the selection list.
↳
The pen symbol indicates a frequency where calibration has not been performed
and where the calibration factor is determined by interpolation. There can be only
one entry with pen symbol at a time. This remains in the list until another fre-
quency value is selected and is then deleted again.
HF probes are broadband probes and cannot measure selectively at the set frequency.
In the case of incorrect frequency specification or multifrequency fields, these settings
can therefore lead to incorrect results.
Non-evaluating probes measure the field strength. The
% of Standard
unit is only avail-
able for selection for these probes if the conversion for a specific personal protection
standard has been activated in the menu. The measured value display in % is related to
the power density and not to the field strength.
Activate conversion:
✓
DEVICE SETTINGS
>
EXPERT SETTINGS
1.
Select
Standard
.
2.
Reference frequency of the field (
Reference Frequency
) indicated.
3.
Activate
% of Standard
: Select
Lowest Limit
or
Reference Frequency
.
The reference to the lowest limit value of a standard (
Lowest Limit
) is recommended
when the frequency of the field source is not known.
The
Lowest Limit
setting is not available if the use of the correction factor is switched
on.
Содержание FieldMan
Страница 1: ...FieldMan Measurement Device for Electromagnetic Fields User manual...
Страница 6: ...6 FieldMan Narda...
Страница 10: ...1 About this product 10 FieldMan Narda...
Страница 14: ...2 General safety instructions 14 FieldMan Narda...
Страница 30: ...5 Controls and display elements 30 FieldMan Narda 5 1 Controls 2 3 4 5 6 1...
Страница 90: ...8 Operating modes 90 FieldMan Narda...
Страница 108: ...10 Cross mode functions 108 FieldMan Narda...
Страница 112: ...11 PC software and updates 112 FieldMan Narda...
Страница 120: ...13 Conformity 120 FieldMan Narda...
Страница 125: ......