63
Chapter 9
Time-Averaged Measurements
To initiate Mode 2 averaging (averaging to meet the
requirements of a specific standard)...
1. Make sure that the meter is set for the correct standard
or change it if needed. See Chapter 11,
Using the
Standards Feature
.
2. Make sure the meter is set to the correct frequency for
the measurement you are making. See page 64.
3. Press number
2
on the keypad to select
STANDARDS
AVG
from the averaging menu screen. Selecting
Mode 2 averaging will automatically shut off Mode 1
averaging if it is on.
4. A new screen will appear stating that averaging has
been enabled in accordance with a specific standard.
5. Press
F3
(
CANCEL
) to return to the initial time
averaging menu screen to make another selection or
press
F4
(
OFF
) to turn time averaging off and return to
the standard measurement screen.
6. Press
F1
(
BEGIN
) to automatically return to the
measurement screen with new options available for the
function keys.
A V E R A G I N G E N A B L E D
S T N D : I E E E C 9 5 . 1 - 1 9 9 9
L E V E L : C O N T R O L L E D
B E G I N C A N C E L O F F
F1
F2
F3
F4
6 . 2 6
W
m
c m
/
2
1 1 . 5 M A X
1 0 . 0 0 G H z
S T A R T M E N U F R E Q Q U I T
F1
F2
F3
F4
Содержание 8718B
Страница 9: ...8 Operating the Meter...
Страница 37: ...36 Operating the Meter...
Страница 71: ...70 Operating the Meter...
Страница 75: ...74 Operating the Meter...
Страница 79: ...78 Operating the Meter...
Страница 87: ...86 Operating the Meter...
Страница 104: ...103 Chapter 18 User s Software Main Menu Screen...
Страница 108: ...107 Chapter 18 User s Software Probe Installation Screen...
Страница 112: ...111 Chapter 18 User s Software Logged Survey Data Screen...
Страница 124: ...123 Appendix B Theory of Operation Model 8718B Functional Block Diagram...
Страница 129: ...128 Appendices...
Страница 133: ...132 Appendices...