General operation
Error code
1.
Check a chip is inside of
device when turning on the
device.
2.
Removing the chip first,
then rebooting the device.
3.
If this message come up
again after removing a chip,
contact your local distributor or
Frames with errors are over
50% of total counting frame.
Frame with error is a frame
that contains cells whose
diameter is larger than 30μm.
Cells are more than 4X10
6
cells/mL.
“Over detection range”
will
be shown on report program.
Cells are more than 2X10
6
cells/mL.
“High concentration cells”
will be shown on report
program.
Cells are less than 4X10
5
cells/mL.
“Low concentration cells”
will be shown on report
program.
Cells are less than 5X10
4
cells/mL.
“Under detection range”
will
be shown on report program.
E :
O :
H :
L :
U :
[ 2 channel ]
[ 4 channel ]