7. Fault alarm information 

The following table lists the alarm information and related fault causes and 

handling methods emitted by the machine. You can follow the prompts to 

troubleshoot; If the user cannot troubleshoot after trying or the alarm 

message sent is not listed below, the user should immediately contact our 

company's maintenance personnel.

 

After an abnormality occurs, the power should be turned 

off first and restarted after the fault is resolved.

 

Table 3: Fault alarm information

 

 

Fault 

codes

 

Code Definition 

Troubleshooting 

E2 

overspeed

 

Speed up too fast, lower the gear up;

 

E3 

Emergency stop switch

 

  System emergency stop;

 

E4

 

Hall fault

 

  Check if there is an open circuit at both ends of the 

outlet wire of the emergency stop switch detection switch 
(the button should be in a rotated out state); 

E5

 

Brake overpressure

 

  The motor Hall detection is incorrect;

 

E7 

No speed measurement

 

  Check if the Hall plug of the motor is properly connected;

 

E8 

Communication 
error

 

  Check if the brake resistor is properly connected or if 

the brake resistor is burnt out; 

E9 

Brake overpressure

 

  Reduce the DEC value; 

Содержание LGB004

Страница 1: ...tube vortex mixer safety rules For safety reasons please read the operating manual carefully before starting the operation Components subjected to heat humidity and biological changes can lead to fun...

Страница 2: ...ters 6 3 Introduction to the appearance and functions of the multi tube vortex mixer 7 8 4 Operating Instructions 8 11 4 1Introduction to the Installation of Test Tube Mats 9 4 2 Turn on off the power...

Страница 3: ...ironmental impacts including natural ultraviolet radiation Corrosion and wear of protective cover parts and other safety parts Indoor use Altitude 2000m The applicable ambient temperature range of the...

Страница 4: ...anual carefully before using the instrument Note Danger of high voltage Danger of electric shock Meaning of safety instructions In order to avoid harm to the surrounding people or damage to surroundin...

Страница 5: ...face When installing the gasket or fixture make sure that it is in place to prevent the gasket and fixture from flying out during work Do not oscillate more than the maximum load of the instrument I P...

Страница 6: ...s drug analysis Drug development and other scientific fields in food safety It shows incomparable advantages in handling a large number of sample mixing saving time and effort Automatic mixing improve...

Страница 7: ...hout the need to adjust cumbersome parameters 6 The perfect optimization design of the body structure ensures sufficient counterweight weight and circumferential eccentric balance operation providing...

Страница 8: ...lation mode circumference Circumferential diameter 3 6mm Time function display Digital timing 1s 999min Continuous Countdown Jog positive timing vein Speed regulation function display Impulse intermit...

Страница 9: ...ube pressing plate B Lock nut C Upper test tube pad D Test tube rack E Lower test tube pad F Test tube fixed disk G Operation panel H Power switch I Power cord socket The above image shows the LGB004...

Страница 10: ...irm key after adjusting the parameters press this key to confirm 6 Pulse mode set for intermittent operation 7 Stop Start key 8 Click the button to momentarily oscillate 9 Speed display area unit r mi...

Страница 11: ...essary to confirm that they are installed in place to prevent the test tube and sponge pad from flying out during operation If an emergency occurs during use immediately press the emergency stop butto...

Страница 12: ...3 When using any test tube pad in LGB004 evenly place the sample to be mixed and switch to continuous mode jog mode or pulse mode 4 Select continuous mode operation press the MENU key and the space on...

Страница 13: ...esolved please turn out the emergency stop button before starting the instrument again 5 Maintenance matters Do not attempt to clean the instrument when the power cord is plugged in or the power switc...

Страница 14: ...sponge or cotton cloth and then use distilled water to clean the detergent Do not sprinkle or spray the rotor with water as the liquid may be carried away and cause corrosion Allow to dry upside down...

Страница 15: ...ontainers with a diameter of no more than 10mm with a total of 54 holes Test tube gasket Standard configurat ion LGB004 2 Max 2500rpm Used to place test tubes and small containers with a diameter of n...

Страница 16: ...estarted after the fault is resolved Table 3 Fault alarm information Fault codes Code Definition Troubleshooting E2 overspeed Speed up too fast lower the gear up E3 Emergency stop switch System emerge...

Страница 17: ...s the requirements of AC100 250V The instrument has high vibration and noise 1 Check if the instrument is placed on a horizontal desktop 2 Check if the test tube pressure plate is tightly pressed 3 Ch...

Страница 18: ...of this user manual The implementation of installation debugging moving positions adding accessories and maintenance work is the responsibility of the production unit or sales company authorized by th...

Страница 19: ...LGB004 3 1 3 Additional test tube pad According to the order contract 4 Certificate of Conformity 1 5 Quality Assurance Page 1 6 Operating Instructions 1 7 Power cord 1 When unpacking please check in...

Страница 20: ...English 20 User s manual for LGB004 Version 1 July 2023 Maintenance records...

Отзывы: