Temperature
slope
correction
It is usually used to correct errors
in high temperature measurement.
PK = 1000 × (Actual value
–
PV) ÷ PV
(-999
~
999)
0
Parameter table 2
Prompt
Name
Function description
(Setting range)
Factory value
Lc-
Password
key
When “Lc=9”, enter the next
parameters.
0
doT
Temperature
decimal
point
0: No decimal point display
1: With decimal point display
(0
~
1)
1
ndT
Timer mode
0: No timer function.
1: Start timing when the temp
reaches the set value.
2: Start timing as soon as the
controller starts working.
(0
~
2)
1
Hn
Timer unit
0: Minute.
1: Hour.
(0
~
1)
0
SPd
Timer
parameter
If “ndT=1”, start timing when
“SV
-
SPd≤PV≤SV
+
SPd”
(0.1
~
50.0
℃
)
0.5
EH
Timer end
mode
0:
Continue
to
control
the
temperature
1: Stop temperature control
(0
~
1)
0
oPn
Door
parameter
Automatic judge door opening.
0: invalid; 0: valid
(0
~
1)
0
nP
Power
percentage
Percentage of max heating power
output.
(0
~
100%) 100
Co
Heating
prohibited
deviation
When “PV≥SV+Co”,
heating output will be cut off
(0
~
50.0
℃
)
50.0
SPH
Max set
value
The maximum temperature set point
value.
(0
~
100.0
℃
)
100.0