Page 27
ENGLISH
Version 1, September 2023
User’s manual for analytical balances
Balance back to zero
External calibration (single-point calibration)
Linear calibration
Enter external calibration mode; “CAL-100” flicker
Put the 100g weight (for example, in balance ref.
KBF010)
Put the 100g weight (for example, in balance ref.
KBF010). Screen shows “CAL-100”. Remove the
weight. Finish the calibration.
Balance back to zero
Press MENU button until screen shows “SET-C”
Enter linear calibration mode; screen shows “Lin-0”
Finish the calibration
Follow the prompts to put the corresponding weights