NAG/IM/HHP 1000/01/22 

11

 

 
 

 

Careful use of the pressure release valve and fine 

adjustment valve enables a controlled release of 

pressure, essential for calibration purposes. 

 

 

Handles are to be fully squeezed together on 

each stroke to achieve maximum pressure 

output. 

 

 

Allow the pressure to settle for few minutes 

before taking the readings, to avoid pressure 

drop due to thermodynamic effects, setting of 

seals and expansion of pressure measuring hose.

 

 

 

As soon as dirt is noticed, change the oil.  

 

 

 

Do not release maximum pressure using Pressure 

release valve, but use fine adjustment valve to 

avoid damage to the instrument. 

 

 

Do not exceed maximum pressure indicated on 

pump label. 

 

 

It is not advisable to use hydraulic oil operated 

pump for water / oxygen ranges by removing oil 

& replacing with distilled water & vice versa. 

Do 

not use oil in water operated hand pump as it 

would damage the inner seals & critical parts.

 

 

 

 

Содержание HHP 700

Страница 1: ...Nagman HYDRAULIC HAND PUMP HHP 700 1000...

Страница 2: ...ince 1972 serving Industries in India Worldwide thro the Manufacture Supply of World Class Calibration Instruments Systems like Temperature Pressure Signal Calibrators Black Body Calibration Sources P...

Страница 3: ...ck to correct any errors noticed and to improve the quality of contents of this Manual Specifications are subject to change owing to continuous development and we reserve rights to effect Changes Modi...

Страница 4: ...No Description Page No 1 Introduction 5 2 Specification 6 3 Standard Delivery Optional Accessories 7 4 Parts Identification 8 5 Connection Diagram 9 6 Safety Instructions 10 7 Operating Instructions 1...

Страница 5: ...rce for calibrating pressure transmitters pressure transducers and pressure gauges There is a Low High pressure selector switch in the pump which is used for selecting low high pressure output It is f...

Страница 6: ...dicator Connection Port 3 8 F BSP Test Connection Port 1 4 F BSP Medium Oil Water Optional Oil Reservoir Material Transparent Acrylic Reservoir Capacity 100 cc Stabilization time 15 20 minutes Instrum...

Страница 7: ...icate Instruction Manual Optional Accessories Carrying Case Soft Hard Analog Digital Pressure Gauges Quick Connector Spare Seals Kit Hydraulic Hose with end fittings Length 1 Meter Water operated Hand...

Страница 8: ...r to Master Instrument 2 Pressure Release Valve 3 Fine Adjustment Valve 4 Front Port 1 4 BSP female 5 Reservoir 6 Reservoir Filling Plug Safety valve 7 Low High Pressure Selector 8 Fluid Inlet tube 9...

Страница 9: ...NAG IM HHP 1000 01 22 9 5 CONNECTION DIAGRAM 1 Master Reference Measurement 2 Unit Under Test...

Страница 10: ...NAG IM HHP 1000 01 22 10...

Страница 11: ...ure drop due to thermodynamic effects setting of seals and expansion of pressure measuring hose As soon as dirt is noticed change the oil Do not release maximum pressure using Pressure release valve b...

Страница 12: ...losed in clockwise direction Operate fine adjustment valve 3 bring it to mid position in anti clockwise direction Fully squeeze handles 10 in and turn Low High pressure selector 7 to low position The...

Страница 13: ...ssure and then change the selector position to high for generating required maximum pressure depends on model Generate the remaining pressure using fine adjustment valve by turning it in clock wise di...

Страница 14: ...pumping Check for any leak at all port connections and ensure for no trace of oil leak at connection ports Check that the Release Valve is fully closed Pump servicing work should be carried out only...

Страница 15: ...nal packing Nagman s liability ceases if Parts are replaced repaired using spare parts which are not identical to those recommended by the manufacturer Nagman s liability is restricted to errors that...