TY-KH15-009A Ver. 3 

 

 

Cautions for proper and safe operation 

To use this product safely, please read this document and the instruction manual carefully. 
Incorrect operation may cause malfunction, damage, or an accident.  
The cautions and warnings in the document are indicated by the following:  
 

 

Indicates that incorrect use may result in severe injury or loss of life. 

 

 

 

Indicates that incorrect use may result in risk of injury or property damage. 

 

 

 

 
 
 
 
 
 

 
 
     

Warni ng 

  Do not apply pressure greater than the maximum allowable pressure.   

This may cause damage or burst of the pressure element and result injury or 
damage to the surroundings. 
 

  Do  not  use  this  product  to  measure  the  object  which  causes  gas/liquid 

contact area to corrode.   

 

This  may  cause  damage  or  burst  of  the  pressure  element  and  release  of  the 
measurement object. It will result in injury or damage to the surroundings.  

 

  Do not apply excessive weight, vibration or shock.  

 This  may  cause  damage  or  burst  of  the  unit  and  release  of  the  measurement 

object. It will result in injury or damage to the surroundings.  

 

  Use of a non designated power source may cause a fire or electrical shock.  

 

  Use this device within the operating temperature range.  

  Using  outside  the  range  of  operating  temperature  may  cause  malfunction  or 

damage of the product and result in injury or damage to surroundings.  

 

  Perform  wire  connection  according  to  the  wire  connection  faceplate  or  the  wire 

connection instructions of the instruction manual. 
Incorrect wire connection may cause injury or a fire.  

 

  Use explosion -proof product where the place with risk of existence of explosive gas. 

It may cause ignition or explosion. 

 

  Where the measurement object is oxygen, use oil-proof product.   

In  case  of  ordinary  product,  there  is  a  possibility  that  oil  remains  inside  the 
product. It may cause ignition or explosion by reaction with oxygen.  

 

  Perform installation carefully according to the installation instructions of the 

instruction manual. 

 

  Do  not  alter  this  product.  Also,  do  not  remodel  this  product  by  adding  new 

functions. 

Contact NKS for repairs. 

 

  Since  this  product  is  a  precision  measuring  instrument,  keep  the  product  as  far  away 

from noise source as possible. 

Also, remove noise from the power source by using noise filter, etc.  

 

  For  applications  where  serious  damage  or  accidents  may  occur  due  to  malfunction  or 

failure of this product, use several units to configure various protection circuits according 
to the degree of seriousness of such damage or accidents, such as "1 out of 2", "2 x (1 out 
of 2)", "2 out of 3", etc. 

 

* If breakdown or malfunction of this product may directly threaten human life or it may 

harm human body, please contact NKS in advance. 

 

Warning 

Caution 

 

Содержание KH15-M Series

Страница 1: ...TY KH15 009A Ver 3 Type KH15 Pressure Transmitter Instruction Manual Jun 2023 8 1...

Страница 2: ...ay cause malfunction or damage of the product and result in injury or damage to surroundings Perform wire connection according to the wire connection faceplate or the wire connection instructions of t...

Страница 3: ...f external connection 5 1 Wire connection 5 2 Examples of external connection 5 7 Operating principle 6 8 Precautions for use 1 Installation 7 2 Cable specifications 7 3 Cautions for transportation an...

Страница 4: ...ressure transmitter with semiconductor strain gauge in the pressure detection part Standard outputs are 4 to 20mADC 0 to 5VDC and 1 to 5VDC The product has an adjustment window on the side surface whi...

Страница 5: ...t and span Where accuracy is 1 0 F S 0 1 F S For both zero point and span Responsiveness 1 ms or shorter Liquid contact material Diaphragm SUS630 17 4PH Fitting SUS316 O ring NBR Standard EPDM Fluorin...

Страница 6: ...TY KH15 009A Ver 3 4 10 5 Outline diagram...

Страница 7: ...r Red Power supply Wire color Red Power supply Wire color White Power supply Wire color White COM Wire color Green Output 2 Examples of external connection e g 1 2 wire system 4 to 20m ADC e g 2 3 wir...

Страница 8: ...the semiconductor strain gauge The detection circuit is the full bridge type whose four sides consist of semiconductor strain gauge From this bridge circuit electric signals proportional to the strai...

Страница 9: ...n the detection sensor part and result in failure Make sure to take countermeasures against condensation 2 Cable specifications Where cable sheath is removed in case of current output 2 wire type ther...

Страница 10: ...powered on before starting main operation 10 Maintenance and adjustment 1 Maintenance This product rarely needs maintenance since it does not have any movable part in the detection part due to its st...

Страница 11: ...a Compound pressure range 5 45mA In case of voltage output 1 to 5 VDC Output in air ventilated status V 4V Upper limit output in the range 0 1MPa x 0 1 1V e g 0 5 MPa Compound pressure range 1 667V 1...

Страница 12: ...harge However note that the following cases are excluded 1 Where the delivered products are disassembled altered or where their parts are replaced or new function is added by the customer or any third...