background image

DATA TRANSMISSION – SERIES RS-232 INTERFACE (only for communication)

1. iBalance 0232C,s -RS232 C’s UART signal
2. Format
(1) Baud rate:  1200 bps  2400 bps   4800bps   9600 bps
(2) Data bits: 

8 bits

(3) Parity bit: 

none

(4) Stop bit: 

1 bit

(5) Code ASCII

DATA FORMAT

HEAD1 HEAD2  DATA 

 

               UNIT        CR

12 3    4 56        7 8 9 10 11 12 13 14 15 16 17   18 19 20    21 22 23 24 25

HEAD1 (2BYTES)   

                 HEAD2 (2BYTES)

OL - overload   

 

NT - net weight mode

ST - stable 

 

 

US - unstable

DATA(8BYTE)

2D  (HEX) = “- ”(negative sign) 

20 (HEX) = “   ”(blank)

2E  (HEX) = “. ”(decimal point)

UNIT (4 byte)

g = 20 (HEX);  20 (HEX); 

20 (HEX); 

67 (HEX)

kg = 20 (HEX);  20 (HEX); 

6B (HEX); 

67 (HEX)

ct = 20 (HEX);  20 (HEX); 

63 (HEX); 

74 (HEX)

ozt = 20 (HEX);  6F (HEX); 

7A (HEX); 

74 (HEX)

CR = OD (HEX);  OA (HEX)

TRANSMISSION EXAMPLE

stable net + 0.168 g
HEAD, HEAD, DATA  UNIT  CR
ST, 

NT 

+ 0.168    g 

OA, OD

PRESET UNITS

ADDITIONAL UNITS

g, ct, oz, ozt, lb, dr, gn, dwt, gsm, PCS , MM, tl.J, tl.T. tl.H,
t, 1/8, 1/4, TMr, TAR

Units

Capacity

SPECIFICATIONS

267mm x 191mm x 70mm
127mm x 146mm

DC 6 V Rechargeable Battery or power adaptor

Optimum 10-40°C (50-104°f)

3100g x 0.01g

Scale dimension
Tray dimension

1100g

Scale Weight

Up to scale’s maximum capacity

Tare range

Operating temperature
Power Source

PRECISION

0.01g

• g

 GRAM

• ct

 CARAT

• oz

 OUNCE

• ozt

  TROY OUNCE

• 

lb 

POUND

• dr

 DRAM

• gn

  GRAIN (UK)

• dwt

  PENNY WEIGHT

 gsm 

GRAMS PER METRIC SQUARED

• MM

  MOMME (JPN)

• tl.J

  JEWELRY TAEL (HONG KONG)

• tl.T

  TAEL (TWN)

• tl.H

  TAEL (HONG KONG)

• t

 

TOLA (INDIA)

• 1/8

  1/8 OUNCE

• 1/4 

  1/4 OUNCE

 TMr

       TOLA, MASHA, RATTI

• 

TAr

  TOLA, ANNIE, RATTI

Содержание iBalance i3100

Страница 1: ...BALANCE ...

Страница 2: ...We design our scales to be as resistant to shock or drops as possible however there is no way for us to protect 100 against load cell or sensor damage Failure to follow these instructions will void your warranty Always allow the scale to acclimate to normal room temperature for at least one hour before use Give your scale sufficient warm up time Usually 30 60 seconds before calibration to give the...

Страница 3: ...deduct the weight of an item or container The symbol will appear and reading will go to zero Press it again to exit the tare mode when empty the tare indication will disappear CAL is used for entering calibration and other function settings SCALE FEATURES DATATRANSMISSION PORT on right side of the scale ADJUSTABLE FEET on bottom of each corner of the scale OPTIONAL CLEARWIND SCREEN AIR BUBBLE LEVE...

Страница 4: ...e load immediately The unit may return to normal operation Remember You can permanently damage the scale and void your warranty by overloading it The LCD will display ERR if the weight placed on the platform is incorrect during calibration CALIBRATION When to calibrate calibration is RARELY required Calibration may be required when the scale is first set up for use or if the scale is moved to a di...

Страница 5: ...go through the self test COMMUNICATION 2 The scale should now read bAUd followed by 12 24 48 or 96 These values represent the baud rate settings 1200 2400 4800 9600 3 Use to scroll through the settings and press to confirm 4 The display will now read Pr Press Sending Co Stable Sending or St Continuous Sending 5 Use to scroll through the settings and then press to confirm BACKLIGHT 6 The display sh...

Страница 6: ...0 HEX 6F HEX 7A HEX 74 HEX CR OD HEX OA HEX TRANSMISSION EXAMPLE stable net 0 168 g HEAD HEAD DATA UNIT CR ST NT 0 168 g OA OD PRESET UNITS ADDITIONAL UNITS g ct oz ozt lb dr gn dwt gsm PCS MM tl J tl T tl H t 1 8 1 4 TMr TAR Units Capacity SPECIFICATIONS 267mm x 191mm x 70mm 127mm x 146mm DC 6 V Rechargeable Battery or power adaptor Optimum 10 40 C 50 104 f 3100gx0 01g Scale dimension Tray dimens...

Отзывы: