II. Solution Mode Selection 

1. 

Press         ,          or        , to select which parameter is having its 
solution type changed. 

2. 

Press and hold         about 3 seconds or until "SEL" appears on the 
display. 

 

3.  

Use the         or         key to select type of solution desired. The  

possible solution types are KCI, NaCl or 442

4. 

Press         to accept new solution type, or to turn on the SALt 
(salinity) mode.

 

CALIBRATION 

CALIBRATION SOLUTIONS 

Mode 

Standard Solution or Buffer 

Conductivity KCl 

KCl: 7000 

µ

TDS 442

 

442

: 3000 PPM 

Displays as 2027 ppm in TDS NaCl or MIN/SALT mode

 

TDS NACL & MIN/SALT

 

pH and FC

E

  

4.0, 7.0 and 10.0 buffer 

Alkalinity 

100 ppm Alkalinity Standard Solution  

Hardness 

200 ppm Hardness Standard Solution  

FC

E

 ORP calibration is electronic and based on pH Cal Results 

I. CONDUCTIVITY, MIN /SALT  OR TDS CALIBRATION 

1. 

Rinse cell cup 3 times with proper fresh, standard solution. 

2. 

Refill cell cup with standard solution. 

3. 

Press        ,        or         , then press         , "CAL" will appear. 

4. 

Press        or        until display agrees with standard solution. 

5. 

Press          to accept value. 

NOTE

: A deviation of more than 10% from standard solution value will 

produce a "FAC" on the display.  Press          to accept factory calibration, 
or clean the cell and recalibrate. 

 II. pH/ CALIBRATION 

1.

  Rinse sensor well 3 times with 7.0 buffer solution. 

 

2. 

Refill sensor well with 7.0 buffer solution. 

3. 

Press        , then        , "CAL", "7" and "BUFFER" will appear on the 
display. 

4. 

Press         or         until display agrees with buffer value. 

5. 

Press          once to accept the value. 

6. 

Rinse 3 times with either ACID (Acd) (pH 1- 6) or BASE (bAS) (pH 8-
14) buffer solution. 

7. 

Refill and repeat steps 4 & 5. 

8. 

Rinse 3 times with opposite buffer solution, refill and repeat steps  
4 & 5 or press         to exit. 

9. 

When calibration is complete, rinse sensor, then refill sensor well with 
Myron L Storage Solution and reinstall cap. 

III. ORP/ FC

E

 CALIBRATION 

  ORP electrodes rarely give false readings without problems in the 

reference electrode.  

  For this reason, and because calibration solutions for ORP are highly 

reactive and potentially hazardous, your PoolPro has an electronic 
ORP calibration.  

  This causes the zero point on the reference electrode to be set 

whenever pH 7 calibration is done.

 

MAINTENANCE

I. BATTERY REPLACEMENT (LOBATT) 

1. 

Clean and dry Instrument THOROUGHLY.  

WARNING:

 

ONLY

 open PS6FC

E

 

 in a clean, dry environment! 

2. 

Remove the four (4) bottom screws. 

3. 

Open instrument CAREFULLY.  

4. 

Carefully detach battery from circuit board.  

5. 

Replace with 9-volt alkaline battery.  

6. 

Replace bottom, ensuring the sealing gasket is installed in the groove 
of the top half of case. 

7.

  Re-install the 4 screws, tightening them evenly and securely. 

II. CAUTIONS 

  Solutions in excess of 71°C/160°F should not be placed in the cell cup 

area; this may cause damage. 

  The pH sensor may fracture if the PS6FC

E

  temperature is allowed to 

go below 0°C/32°F. 

  The cell cup should be kept as clean as possible. Flushing with clean 

water following use will prevent buildup on electrodes. If very dirty 
samples  (particularly scaling types) are allowed to dry in the cell cup, 
a film will form reducing accuracy.  

  The pH sensor in your PS6FC

E

 

 should not be allowed to dry out.  

  Whenever the pH sensor is not in use make sure it is filled with  

Myron L

®

 Company Storage Solution and that the protective cap is in 

place. 

III. CLEANING SENSORS 

A.  Conductivity or TDS 

  When there are visible films of oil, dirt, or scale in the cell cup or on 

the electrodes or if readings are not as expected, use isopropyl 
alcohol or a foaming non-abrasive household cleaner. 

  Rinse out the cleaner, and your PS6FC

E

 

 is again ready for accurate 

calibration then measurements. 

B. 

If the pH sensor dries out or becomes dirty, it may be cleaned and/or 
reconditioned.  For instructions on preforming these operations, 
download the full PS6FC

E

 

 Operation Manual from the Myron L

®

 

Company website. 

IV. pH Sensor Replacement 

Order model RPR. Be sure to include the model and serial number of your 
instrument to ensure receipt of the proper type.  

NOTE

: Complete installation instructions are provided with each 

replacement sensor.

 

SPECIFICATIONS 

Spec 

pH 

CONDUCTIVITY 

TDS & MIN / SALT 

ORP 

FC

E

 

TEMP 

Ranges 

0-14 pH 

0 - 9999 

µ

S/cm; 10 - 200 mS  

In 5 autoranges

 

0 - 9999 ppm; 10 - 200 ppt  

In 5 autoranges

 

± 999 mV 

0.00 – 9.99 ppm 

ORP = 350mV - 725 mV, <9.9 pH 

ORP = 725 mV - 825 mV, <8.9 pH 

0°C – 71°C 

32°F – 160°F 

Resolution 

0.01 pH 

0.01 < 100 

µ

0.1  (<1000

µ

S) 

1.0 (<10 mS 

0.01 < 100 mS) 

0.1 mS (

 100 & < 200 mS) 

0.01 < 100 ppm 

0.1 (<1000 ppm) 

1.0 (<10 ppt 

0.01 < 100 ppt) 

0.1 (

 100 & < 200 ppt) 

±.1 mV 

0.01 ppm 

0.1 °C / °F 

Accuracy 

± 0.01 pH 

±1% of reading 

±.1 mV 

±. 0.3 ppm <1.0 ppm;    

±. 0.2 ppm 

1.0 ppm 

±0.1 °C / °F 

COND/TDS 

Ratios 

 

Programmed: KCl, NaCl or 442™          Adjustable 0.20 -07.99 

 

 

 

Temp Co 

Auto: 0-71°C, 32-160°F          Adjustable (COND & TDS)  0 – 9.99% / °C 

 

 

 

 