TECHNICAL MANUAL -
hyperion X5
EN
CALIBRATION
33
Select the “not centered” option in the calibration
tool, then presso “OK” to start a new acquisition in
order to verify the change made.
Repeat the procedure until the x-ray beam is
correctly aligned.
Centering the collimator vertically
Slightly loosen the “A” screws shown in the
following picture, loosen or tighten the “B” screw
to move the collimator up or down. Tighten the “A”
screws before taking the next image.
Select the “not centered” option in the calibration
tool, then presso “OK” to start a new acquisition in
order to verify the change made.
Repeat the procedure until the x-ray beam is
correctly centered.
When the x-ray beam is centered, select
“Centered” and click “OK”.
Содержание hyperion X5
Страница 1: ...hyperion X5 technical manual Cod 97070183 Rev 00 05 2015 www my ray com EN ...
Страница 2: ......
Страница 3: ...EN TECHNICAL MANUAL hyperion X5 Code Rev Date 97071183 00 15 05 ...
Страница 4: ......
Страница 6: ......
Страница 15: ...TECHNICAL MANUAL hyperion X5 EN PC CONFIGURATION 13 Select I accept then click Next Click on Install Wait ...
Страница 44: ...TECHNICAL MANUAL hyperion X5 42 CONTROL BOARDS DESCRIPTION EN 3 2 POWER BOARD 97661407 ...
Страница 45: ...TECHNICAL MANUAL hyperion X5 EN CONTROL BOARDS DESCRIPTION 43 ...
Страница 48: ...TECHNICAL MANUAL hyperion X5 46 CONTROL BOARDS DESCRIPTION EN 3 3 LOGIC BOARD 97661408 ...
Страница 49: ...TECHNICAL MANUAL hyperion X5 EN CONTROL BOARDS DESCRIPTION 47 ...
Страница 52: ...TECHNICAL MANUAL hyperion X5 50 CONTROL BOARDS DESCRIPTION EN 3 4 SENSOR INTERFACE BOARD 97661485 ...
Страница 53: ...TECHNICAL MANUAL hyperion X5 EN CONTROL BOARDS DESCRIPTION 51 ...