47
EN
hyperion
X5 2D/3D
- TECHNICAL MANUAL
TROUBLESHOOTING
6. TROUBLESHOOTING
6.1 CRANIOSTAT ADJUSTMENT
Should the craniostat not be centred with respect to the sagittal laser, perform a mechanical
adjustment:
Proceed as follows:
1.
Check that craniostat is centred
2.
Close the 2 side tabs
3.
Remove lower cover. There are no screws, just pull the cover downwards.
4.
Loosen the 2 screws shown in the figure and move the craniostat to align it to the sagittal laser.
Содержание Hyperion X5 2D
Страница 1: ...www my ray com Cod 97071218 Rev 01 01 2017 hyperion X5 2D 3D technical manual EN ...
Страница 2: ......
Страница 4: ...2 EN hyperion X5 2D 3D TECHNICAL MANUAL CONTENTS ...
Страница 37: ...35 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION ...
Страница 40: ...38 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION PIN 4 signal X ray ON ...
Страница 41: ...39 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION ...
Страница 44: ...42 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION 5 4 SENSOR INTERFACE BOARD 97661485 ...
Страница 45: ...43 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION ...
Страница 47: ...45 EN hyperion X5 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION 5 5 MONOBLOC AREA ...
Страница 51: ......
Страница 52: ......