11. Use device or train device operators in strict accordance with instructions in the
user manual.
12. When taking other weight-loss drugs, it is recommended to stop taking them for
1 to 2 months before treatment. Extension of the course of treatment is
recommended if you wish to lose weight immediately.
13. Don't use the instrument with an empty stomach. After a full meal, wait at least 1
hour before a course of weight loss treatment.
14. When operating, the instrument should fully touch the skin to avoid uneven heat.
15. Start from the lowest energy level and slowly add up.
16. When using this device, the operating parts must be kept moist and dry skin
should be avoided.
17. Clean the instrument with normal saline after operation to ensure its cleanliness
and hygiene and prolong its service life.
18. Apply gel before for RF treatment and always keep it moist during the process.
19. Adjust suction level to whatever you’re most comfortable with during vacuum&RF
treatment to avoid discomfort.
20. Slight red or purple marks may appear after vacuum&RF treatment, which is
normal so there’s need to worry.
5. Troubleshooting & Solutions
1. The instrument cannot be started, but the key light on the back of the instrument
does not work?
A. Make sure the power cord is connected to a valid power socket.
B. Whether the fuse tube on the back of the instrument is loose or burnt out.
2. No RF output of the instrument?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether the treatment area has been cleaned. Grease or oil essential
oil products may cause poor contact between the treatment handle of the instrument
and human body, resulting in no RF output.
3. Weakened RF output?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether non-conductive grease and other substances are stained on
the treatment handle, which will cause poor contact and weaken the output.
C. Please check whether the product used is the adaptive product specified by the
instrument.
4. No suction or very little suction?