PLEASE NOTICE 

Notices about Usage of 808nm Diode Laser Hair Removal Machine 

 

1. You must add distilled water before you turn on machine. 

2. Test machine 
3. When you change all water, please also use distilled water   
4. Must immit DISTILLED WATER. Change all water 1time/month. 
5. Please start to adjust parameters and treat patients immediately after the laser 
be turned on, it is not allowed to keep it in “Standby” status for long time; If not use 
machine, turn off machine will extend life time of handpiece. 
 
 
 
 
 
 
 
 
 
 

Содержание HR-F16

Страница 1: ...achine 2 Test machine 3 When you change all water please also use distilled water 4 Must immit DISTILLED WATER Change all water 1time month 5 Please start to adjust parameters and treat patients immed...

Страница 2: ...r 5 3 1 Device Structure 5 3 2 Application 5 Chapter 4 Working Conditions 6 Chapter 5 Installation and debugging 6 Chapter 6 Technical Specifications 7 Chapter 7 Operation 7 7 1 Starting 8 7 2 Paramet...

Страница 3: ...alcohol ether oxygen and light decomposition such as drug products should be far away peak power laser light radiation 2 1 4 Wrist watch necklace bracelets and other metal jewelry reflecting light are...

Страница 4: ...lectric shock 2 4 System Device Safety 2 4 1 Water current device when laser emits there is high density heat in the hand piece So it requires constant cooling water conduction cooling The machine is...

Страница 5: ...nal standards connections for power 4 4 The power should be well earthed in accordance with the electrical safety regulations in force 4 5 The machine adaptation voltage range is 200 240V 100 120V If...

Страница 6: ...3 The water must be empty during the long term stop using or long distance transportation Change water We advice the machine should change distilled water in one month after using 4 After Loosing the...

Страница 7: ...7 Chapter 7 Operation 7 1 The Starting interface 7 2 Click diode laser to choose the Diode Laser treatment interface...

Страница 8: ...second Frequency recommendation 4 to 6 hertz Refrigeration Control the skin cooling turn red to cooling turn green means no cooling Click ready to work click standby to finish treatment 7 3 The wavele...

Страница 9: ...D After adjustment have an experiment on normal skin to see whether there is change and ask for the patient s feelings Build the data for the clients about the therapy E Choose the skin type F Shootin...

Страница 10: ...e face appears light red The lesion feels hotness and sting while the root of hair has a red spot The symptoms can be relieved after cold spray or cold compress At the same time Scabbing and the color...

Страница 11: ...ing A The working and treatment principle of the machine B Operation procedure C Common malfunction recognition Don t charge any other costs Please contact our After service department if you had any...