15. Hydro facial, clod hammer, and Ultrasound probes can be used separately and also
together, and the later comes with a better effect.
16. Clean the instrument with normal saline after operation to ensure its cleanliness
and hygiene to prolong its service life.
17. The front of the cleaning scrubber is for deep cleaning and the back is for skin
lifting and nutrients absorption.
18. Various types of hydro facial filters can be cleaned with water, the soaked in 75%
alcohol.
19. Clean the waste bottle immediately after hydro facial treatment.
20. Screw head cleaning: basic cleaning must be done to maintain the prevention of
infection. After using the screw head, disposable toothbrush can be used to brush out
the dirty stuff in the groove of the screw head, rinse tit off with clean water, then soak
with alcohol for 3 minutes, and finally rinse the alcohol ingredients.
21. Internal cleaning (after each bottle of solution is used):
Replace the probe with a cleaning head, fill the solution bottle with half a bottle of
purified water, turn on the device, plug the probe by hand, adjust the water flow to
the maximum, and let the water flow back to clean the residual solution in the pipe.
Around a minute.
22. All parts of the instrument must be tightened, or internal leakage will reduce the
air pressure, resulting in no suction or small stream of water.
5. Troubleshooting & Solutions
1. The instrument can't start and the key lamp on the back of the instrument doesn't
work?
A. Ensure that the power cord is connected to a working power outlet.
B. Ensure the fuse tube on the back of the instrument is not loose or being burnt.
2. No RF output of the instrument?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether the treatment area has been cleaned. Grease or oil essential
oil products may cause poor contact between the treatment handle of the instrument
and human body, resulting in no RF output.
3. Weakened RF output?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether non-conductive grease and other substances are stained on
the treatment handle, which will cause poor contact and weaken the output.
C. Please check whether the product used is the adaptive product specified by the
instrument.