CALIBRATION 

When to calibrate 

Calibration may be required when it is initially installed, if the scale/balance is 
moved to a substantial distance. 
This is necessary because the weight of a mass in one location is not necessarily the 
same in another location. Also, with time and use, mechanical deviations may 
occur. 
 

How to calibrate 

 

1.

 

Before entering calibration mode, press 

[ON/OFF]

 to turn the power on for 

1minutes.

 

2.

 

Enter calibration mode 

3.

 

The scale must be powered OFF

.  Press and hold 

[ZERO]

 first, then Press 

 

[ON/OFF]

 while keeping the 

[ZERO] 

button depressed, then release both keys. 

The display will show the A/D value (a series of random numbers). 

4.

 

Wait for the stable indicator to be displayed, Press 

[UNIT]

 the display will show 

"00SAVE"

, then after 2 seconds the display will return to the random A/D value. 

5.

 

Place

 50

00 grams of weight(s) on platform.  Wait for the stable indicator to 

display, then press 

[UNIT]

, the display will show 

"00SAVE"

.  Calibration is 

complete, remove the weight(s).

 

Turn the scale OFF (Press the ON/OFF key) , then turn it back ON and check some 
weight readings.    If calibration is still incorrect, repeat calibration but try it more 
slowly 

FEATURES 

z

 

Auto shut off (optional).            

   

   

Auto calibration 

z

 

Auto zero tracking                                 

   

 

Auto backlight (optional) 

z

 

Low batter indication               

 

g/oz/lb/ozt/gn/dwt/ct/n or   

z

 

Large LCD (6digits 15 mm high)         

   

  g/ct/oz/tl/tola/gn/dwt 

z

 

Large square weighting pan             conversion(optional) 

z

 

Net weight/stability indication        

 

with counting function(optional) 

z

 

1/30000&1/15000 division available 

PARTS DESCRIPTION 

 

RIGHT SIDE 

         

AC adaptor 

Socket 

 

 TOP 

 

Weighing platform 

The arch face of platform is front.         

Correct the direction of the pan. 

 

 

 

 

Display 

 

ON/OFF

 key 

Turns the scale 

 

power on/off. 

                                                                                                PCS key 

UNIT 

key                                      counting mode 

Changes weighing units, g ,oz,lb, ozt, dwt, ct,gn or pcs(counting mode). 

 

ZERO 

key (for TARE) 

ets display to zero or   

Subtracts weight of a container. 

 

 BOTTOM 

 

Battery compartment 

 

Battery compartment cover     

 

Batteries  

Use 6 AA batteries 

 

DISPLAY  

STABLE

 indicator                     Auto backlight function indicator 

Indicates when the reading is stable.       

                                                                                   

Auto shut off function indicator 

                                                                               

                                                                                                                                                                   

ZERO

 indicator 

                                                                              Indicates when the reading is stable

 

Weighing unit

 

   

                                                                         

 

  Low battery indicator

 

Содержание iBalance i5500

Страница 1: ...cedure If necessary press ZERO key to set the display to 0 3 Place a given number of samples of an item on the pan the Sample Size should be either 10 20 50 or 100 pieces The weight of these samples will show on the display 4 Press the UNIT key several times to put the scale in PCS mode 5 Select the sample size the same as you chose above by pressing the PCS key press it as many times as necessary...

Страница 2: ...causes damage if inside of the scale balance is wet Matter charged static electricity could influence the weighing Discharge the static electricity As example the method is to use the electrification prevention spray to spray it to both side of weighing platform 1 SPECIFICATIONS Model number i5500 Capacity 100g Division 0 005g Platter Tray 140 165mm Net gross weight 510 620g Standard carton 29 22 ...

Страница 3: ... the scale OFF Press the ON OFF key then turn it back ON and check some weight readings If calibration is still incorrect repeat calibration but try it more slowly FEATURES z Auto shut off optional Auto calibration z Auto zero tracking Auto backlight optional z Low batter indication g oz lb ozt gn dwt ct n or z Large LCD 6digits 15 mm high g ct oz tl tola gn dwt z Large square weighting pan conver...

Страница 4: ... possible please allow the balance to warm up for 10 minutes after first turning to power on so that the balance will function Properly and accurately Error messages EEEE Overload Low battery Calibration may be required before weighing Read CALIBRATION first and if necessary calibrate your scale balance for accurate weighing Weighing procedures 1 Press ON OFF to turn on the scale balance When powe...

Отзывы: