Resistance measurement 

Range

Intrinsic 

uncertainty

Observation

200.0 

0.5 %+3 ct

No-load 
voltage: 

approx. 0.4V

2.000 k

0.5 %+2 ct

20.00 k

200.0 k

2.000 M

20.00 M

1.5 %+3 ct

Continuity measurement

Range

Intrinsic 

uncertainty

Observations

200.0 

0.5 % + 3 ct

- No-load voltage: 

approx. 0.4V

 

 : R < 90 

 ± 40 

ENVIRONMENTAL CONDITIONS
Indoor use
Altitude < 2.000 m
Degree of pollution: 2

In use

In storage

Temperature

-10 °C ... +50 °C

-20 °C ... +60 °C

Relative 
humidity

≤ 

80 %RH 

(without 

condensation)

≤ 

70 %RH 

(without 

condensation)

CONSTRUCTIVE CHARACTERISTICS 
Dimensions L x W x H: 181 x 92 x 57 mm
Weight: approx. 400 g

POWER SUPPLY 
Battery: 2x1.5V AA/LR6
Mean battery life: ~600 hours
Auto power off delay: After 15 minutes without action on 
the keys and/or on the switch.

COMPLIANCE WITH INTERNATIONAL STANDARDS  
Compliant with standards IEC 61010-1 and IEC 61010-
2-033 for 600V CAT III installations.
Maximum input voltage: 600V between terminals.

ELECTROMAGNETIC COMPATIBILITY
Emissions and immunity in an industrial environment 
per EN 61326-1.

MAINTENANCE

Except for the fuse and the batteries, the 

instrument contains no parts that can be replaced by 
personnel who have not been specially trained and 
accredited. Any unauthorized repair or replacement of a 
part by an “equivalent” may gravely impair safety.

CLEANING
Disconnect the unit completely and turn the rotary switch 
to OFF.

Use a soft cloth, dampened with soapy water. Rinse with 
a damp cloth and dry rapidly with a dry cloth or forced 
air. Do not use alcohol, solvents, or hydrocarbons.

REPLACEMENT OF THE BATTERIES
The 

 symbol indicates that the batteries are dead and 

must be replaced.
To replace the batteries, proceed as follows:

 

„

Disconnect the unit completely and turn the rotary 
switch to OFF;

 

„

Remove the jacket;

 

„

Unscrew and remove the 4 battery compartment 
cover screws;

 

„

Remove the old batteries and insert the new ones, 
with the polarities as indicated.  

Spent batteries must not be treated as ordinary 
household waste. Take them to the appropriate 
recycling collection point.

REPLACING THE FUSE
When the measurement current exceeds the current 
rating of the fuse, the protection fuse may blow.

For safety reasons this fuse must always be 

replaced by an identical model: 
F1: 10x38, type FF, 10A/600V 
F2: 6.3x32, type FF, 630mA/600V 

When the current measured exceeds the rating of the 
fuse, the fuse may blow.

REPAIR
Return the instrument to your distributor for any work to 
be done, whether under the warranty or not.
If you have to ship the instrument, it is best to use its 
original packaging and to state as clearly as possible, 
in a note attached to the equipment, the reasons for the 
transfer.

WARRANTY

The equipment is warranted against defects of materials 
or workmanship, in accordance with the general terms 
of sale. 
During the warranty period (1 year), the instrument must 
be repaired only by the manufacturer, who reserves 
the right to choose between repairing it or replacing it, 
entirely or partially.
If the equipment is sent back to the manufacturer, 
carriage is paid by the customer.

The warranty does not apply in the following cases:

 

„

Inappropriate use of the equipment or use with 
incompatible equipment;

 

„

Modifications made to the equipment without the 
explicit permission of the manufacturer’s technical 
staff;

 

„

Work done on the device by a person not approved 
by the manufacturer;

 

„

Adaptation to a particular application not anticipated 
in the definition of the equipment or not indicated in 
the user’s manual;

 

„

Damage caused by shocks, falls, or floods.

 CHARACTERISTICS

ELECTRICAL CHARACTERISTIC

Reference conditions

Quantity of influence

Reference value

Temperature

23 ± 2°C

Relative humidity

45 to 75% RH

DC measurement

Without AC component

AC measurement

Sine wave without DC

Frequency measurement Square wave without DC

The intrinsic uncertainties are stated in 
± (x % of r Y points) from 10 to 100% of the range

DC voltage measurement

Range

Intrinsic 

uncertainty

Input resistance

200.0 mV

0.5 %+3 ct

~ 100 M

2.000 V

~ 11 M

20.00 V

~ 10 M

200.0 V

600 V

AC voltage measurement

Range

Frequency 

range

Intrinsic 

uncertainty

Input 

resistance

200.0 mV

40 Hz - 
500 Hz

1 %+10 ct

~ 11 M

2.000 V

1 %+5 ct

~ 10 M

20.00 V
200.0 V

600 V

DC current measurement
Intrinsic uncertainty: 1.5% + 3 ct

Range

Protection

200.0 μA

Fuse 0,63 A / 600 V

2000 μA

20.00 mA
200.0 mA

2.000 / 10.00 A*

Fuse 10 A / 600 V

* 20 A for 30 s.

AC current measurement
Intrinsic uncertainty: 1.5% + 5 ct

Range

Frequency

Protection

200.0 μA

40 Hz - 500 Hz

Fuse 0,63 A

/ 600 V

2000 μA

20.00 mA
200.0 mA

2.000 

/ 10.00 A*

Fuse 10 A / 

600 V

* 20 A for 30 s.

Diode test  

Range

Intrinsic uncertainty

No-load 

voltage

2.000 V

10 %

~ 1.5 V

 

FF 630mA/1000V

FF 10A/1000V

F1

F2

11 - 2013

694308A02 - Ed. 2

http://www.chauvin-arnoux.com

TO ORDER

DMM 111 ... ................................................ P06231420Z