8. Live wire (Figure 3)
1) Switch function range to LIVE scale.
2) Plug the red probe Into “VΩ” socket, unplug the black probe from the socket (suspended),use the red probe to touch the
socket or exposed wire to distinguish live wire and neutral wire.
3) If neutral wire or uncharged object is detected, “- - - -” keep unchanged.
4) When it sense a live wire with a AC voltage over 70V,the LCD display “LIVE” and sound-light LED indication appears.
Attention:
• When measuring LIVE Function,please unplug the black probe from COM terminal to avoid that COM input terminal
interferes the accuracy of distinguishing the neutral/live wire.
• If there is high voltage/current,the meter may provide an inaccurate measurement result. An accurate measurement result
is based on the LCD display and beep frequency.
10. Technical Index
Accuracy: ±(% r digit number), periodical calibration is one year. Temperature and humidity: 23°C ±5°C ≤%80RH.
Temperature coefficient: On the condition of 18°C to 28°C, environment temperature fluctuation is within ±1°C. When the
temperature is < 18°C or > 28°C, add a temperature coefficient error: 0.1 × (specified accuracy)/ °C.
(1) DC voltage
Range
Resolution
Accuracy
Overload protection
4.000V
0.001V
± (0.5%+3)
600 Vrms
40.00V
0.01V
400.0V
0.1V
600V
1V
• Input impedance≥toMO.
• The identified minimum voltage is about 0.5V.
• Accuracy scope: 1% to 100% range (manual mode).
(2) AC voltage
Range
Resolution
Accuracy
Overload protection
4.000V
0.001V
±(1.0%+5)
600 Vrms
40.00V
0.01V
±(0.8%+3)
400.0V
0.1V
600V
1V