8. Live wire (Figure 3)

1) Switch function range to LIVE scale.
2)  Plug the red probe Into “VΩ” socket, unplug the black probe from the socket (suspended),use the red probe to touch the 

socket or exposed wire to distinguish live wire and neutral wire.

3) If neutral wire or uncharged object is detected, “- - - -” keep unchanged.
4) When it sense a live wire with a AC voltage over 70V,the LCD display “LIVE” and sound-light LED indication appears.

 Attention:

•   When measuring LIVE Function,please unplug the black probe from COM terminal to avoid that COM input terminal 

 

interferes the accuracy of distinguishing the neutral/live wire.

•   If there is high voltage/current,the meter may provide an inaccurate measurement result. An accurate measurement result 

is based on the LCD display and beep frequency.

10. Technical Index

Accuracy: ±(% r digit number), periodical calibration is one year. Temperature and humidity: 23°C ±5°C ≤%80RH.
Temperature coefficient: On the condition of 18°C to 28°C, environment temperature fluctuation is within ±1°C. When the 

temperature is < 18°C or > 28°C, add a temperature coefficient error: 0.1 × (specified accuracy)/ °C.

(1) DC voltage

Range

Resolution

Accuracy

Overload protection

4.000V

0.001V

± (0.5%+3)

600 Vrms

40.00V

0.01V

400.0V

0.1V

600V

1V

•  Input impedance≥toMO.
•  The identified minimum voltage is about 0.5V.
•  Accuracy scope: 1% to 100% range (manual mode).

(2) AC voltage

Range

Resolution

Accuracy

Overload protection

4.000V

0.001V

±(1.0%+5)

600 Vrms

40.00V

0.01V

±(0.8%+3)

400.0V

0.1V

600V

1V

Содержание MP730835

Страница 1: ...Page 1 V1 0 24 08 21 Newark com multicomp pro Farnell com multicomp pro Element14 com multicomp pro Smart Digital Multimeter User Manual Part Number MP730835 ...

Страница 2: ...61010 2 033 EN61326 1 safety standard of electromagnetic radiation protection and meet with safety standard of double insulation over voltage CAT Ill 600V pollution level 2 If user do not operate according to the instructions the protection provided by the meter may be impaired 1 Before use please check the meter and probe to make sure there is no any damage or abnormity If the meter or probe is s...

Страница 3: ...ty Overload display OL or OL Drop proof one meter Power supply 2 AAA 1 5V batteries Dimension 130mm 65mm 2Bmm Weight About 137g including batteries Altitude 2000 meters Operating temperature and humidity 0 C to 30 C 80 RH 30 C to 40 C 75 RH 40 C to 50 C 45 RH Storage temperature and humidity 20 C to 60 C SQ RH EMC RF field 1V m Overall accuracy specified accuracy 5 range RF field 1V m No specified...

Страница 4: ... loose Do not input AC DC voltage over 30V for safety 3 Continuity detection 1 Plug the red probe into VΩ socket plug the black probe into COM socket 2 Scan intelligent identification state can be set if users want to switch to manual mode select range scale and connect the probe to the terminals of circuit load in parallel If the resistance is 300 the buzzer beeps continuously which means the cir...

Страница 5: ...el When the measured current is greater than 5A the measurement should be less than 10s the measured interval should be more than 5 minutes 7 NCV Figure 2 Electric field detection When the NCV sensing end gets close to the charged electric field such as the socket or the insulated wire the LCD displays or the buzzer beeps green LED flicker As the electric field strength increases the LCD display m...

Страница 6: ...shing the neutral live wire If there is high voltage current the meter may provide an inaccurate measurement result An accurate measurement result is based on the LCD display and beep frequency 10 Technical Index Accuracy reading digit number periodical calibration is one year Temperature and humidity 23 C 5 C 80RH Temperature coefficient On the condition of 18 C to 28 C environment temperature fl...

Страница 7: ... 999A 0 001A The identified minimum current is about 1mA Accuracy scope 5 to 100 range 5 Continuity test Range Resolution Accuracy Overload protection 400 00 0 10 300 buzzer beeps 50Ω buzzer not beep the open circuit voltage is about 2 0V 600 Vrms 6 Resistance Range Resolution Accuracy Overload protection 400 0Ω 0 1Ω 1 0 2 600Vrms 4 000kΩ 0 001kΩ 0 8 2 40 00kΩ 0 01kΩ 400 0kΩ 0 1kΩ 4 000MΩ 0 001MΩ ...

Страница 8: ...xposed wire 70V AC 50Hz 60Hz 1 If not detected and AC symbol Is displayed 2 If neutral wire is detected keeps unchanged 3 When Live wire of electric supply is detected LIVE and is displayed on LCD the beep frequency and LED flickering frequency will change based on the sensing strength Red light indication LIVE flickers and buzzer beeps 11 Maintenance and repair Figure 4a Attention Please make sur...

Страница 9: ...Information and the suitability of the products for their purpose and not make any assumptions based on information included or omitted Liability for loss or damage resulting from any reliance on the Information or use of it including liability resulting from negligence or where the Group was aware of the possibility of such loss or damage arising is excluded This will not operate to limit or rest...

Отзывы: