Models MP720644 and MP720774 

 

Quick Guide 

 

Содержание MP720644

Страница 1: ...Models MP720644 and MP720774 Quick Guide...

Страница 2: ...ficient 11 How to Use the Probe Safely 11 How to Implement Self calibration 12 Introduction to the Vertical System 12 Introduction to the Horizontal System 14 Introduction to the Trigger System 14 Tou...

Страница 3: ...rminal When powered by AC power it is not allowed to measure AC power source directly because the testing ground and power cord ground conductor are connected together otherwise it will cause short ci...

Страница 4: ...ble potentially Caution It indicates a potential damage to the instrument or other property might occur Safety Symbols Symbols on the product The following symbol may appear on the product Hazardous V...

Страница 5: ...Warning To avoid fire or electrical shock when the oscilloscope input signal connected is more than 42V peak 30Vrms or on circuits of more than 4800VA please take note of below items Only use accessor...

Страница 6: ...3 4 7 12 11 9 6 2 5 8 13 Figure 3 1 Front panel 1 LCD touchscreen 2 CH1 CH2 buttons Access to setting menu of CH1 CH2 3 Power button Power on off 4 The A knob When one of the CH1 CH2buttons is lighte...

Страница 7: ...ngs 10 Autoset button Quick way to apply a set of pre set functions to the incoming signal and display the best possible viewing waveform of the signal 11 Single button Set the trigger mode as single...

Страница 8: ...veform to USB flash disk needs to use this port 5 LAN port the network port which can be used to connect with PC 6 Connector for the DC power source via an AC DC adapter Top Panel Channel input Multim...

Страница 9: ...wo blue dotted lines indicate the horizontal position of cursor measurement 18 The waveform of CH2 19 The frequency of the trigger signal 20 The icon shows the selected trigger type e g represents tri...

Страница 10: ...If it is found that the packaging carton or the foamed plastic protection cushion has suffered serious damage do not throw it away first till the complete device and its accessories succeed in the ele...

Страница 11: ...ent set value of the probe on the menu is 10X 2 Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel Align the slot in the probe with the plug in the CH1 conne...

Страница 12: ...match the probe with the input channel The probe which is not compensated or presents a compensation deviation will result in the measuring error or mistake For adjusting the probe compensation pleas...

Страница 13: ...mpensator and connect the reference wire clamp with the ground wire connector of the probe connector and then push the Autoset button on the front panel 2 Check the displayed waveforms and regulate th...

Страница 14: ...efore it is changed again Caution The default attenuation coefficient of the probe on the instrument is preset to 10X Make sure that the set value of the attenuation switch in the probe is the same as...

Страница 15: ...never the change of ambient temperature is 5 or over Before performing a self calibration disconnect all probes or wires from the input connector Click the icon select the Utility softkey select Funct...

Страница 16: ...e difference between the wave form and the signal ground If the channel is under the AC mode the DC component would be filtered out This mode helps you display the AC component of the signal with a hi...

Страница 17: ...it can be found that the Horizontal Time Base displayed in the status bar changes accordingly 3 When the HOR button is lit use the A knob to adjust the horizontal position of the signal in the wavefo...

Страница 18: ...e left top of the display area to force a trigger signal which is mainly applied to the Normal and Single trigger modes Touchscreen Controls You can control the oscilloscope by different gestures The...

Страница 19: ...enu pane Click the icon on the right bottom of the display area a menu will be shown Click to enter the corresponding function menu When OFF is shown click to turn on the function and enter the menu W...

Страница 20: ...e the icon to hide it Note Swipe up down or left right in this area you can make the icon appear and control it When the P icon appears in the full screen swipe left right to control the horizontal po...

Страница 21: ...l screen swipe up down to control the trigger level of the source in the trigger menu Click the L icon to fine turn long press to adjust continuously Control the trigger level of the source in the tri...

Страница 22: ...rol the time base Control the voltage division In the full screen pinch and spread horizontally to change the time base pinch and spread vertically to change the voltage division of the selected chann...

Страница 23: ...the right bottom of the display area click Measure to display the menu for the settings of the Automatic Measurements At most 8 types of measurements could be displayed on the bottom left of the scre...

Страница 24: ...screen automatically How to Use Autoscale This is a very useful function for first time users to carry out a simple and quick test on the input signal The function is applied to follow up signals aut...

Страница 25: ...choose from Using USB Port 1 Connection Use a USB data cable to connect the USB Device port in the right panel of the oscilloscope to the USB port of a PC 2 Install the driver Run the Oscilloscope com...

Страница 26: ...re we set the IP address to 192 168 1 71 Figure 4 2 Set the network parameters of the computer 3 Set the network parameters of the Oscilloscope Software Run the software on the computer choose the Por...

Страница 27: ...lect Save set in the bottom menu it prompts Reset to update the config After resetting the oscilloscope if you can get data normally in the oscilloscope software the connection is successful Figure 4...

Страница 28: ...oscope Software Run the software on the computer choose the Ports settings of the Communications menu item Set Connect using to LAN About the IP the first three bytes is same as the IP in the step 2 t...

Страница 29: ...rts settings in the software in step 3 The Gateway and Subnet mask should be set according to the router Select Save set in the bottom menu it prompts Reset to update the config After resetting the os...

Страница 30: ...Voltage ACV DCV Measuring AC voltage Measuring DC voltage R C Measuring resistance Testing diode Testing for continuity Measuring capacitance Hold ON OFF Freeze the display during measurement Configur...

Страница 31: ...ent V Voltage measurement R Resistance measurement Diode measurement Continuity test C Capacitance measurement 3 Range 4 Measurement display OL is short for overload indicates the reading exceeds the...

Страница 32: ...as often as operating conditions require To clean the instrument exterior perform the following steps 1 Wipe the dust from the instrument and probe surface with a soft cloth Avoid scratching the prot...

Страница 33: ...efer to the User Manual You can get this manual from the supplied CD INFORMATION ON WASTE DISPOSAL FOR CONSUMERS OF ELECTRICAL ELECTRONIC EQUIPMENT When this product has reached the end of its life it...