User Manual
25
Single Trigger
Single trigger has two types: edge trigger, video trigger.
Edge Trigger
: It occurs when the trigger input passes through a specified voltage level
with the specified slope.
Video Trigger
: Trigger on fields or lines for standard video signal.
The two trigger modes in Single Trigger are described respectively as follows:
1. Edge Trigger
An edge trigger occurs on trigger level value of the specified edge of input signal. Select
Edge trigger mode to trigger on rising edge or falling edge.
Push the
Trigger Menu
button to display the Trigger
menu on the right. Select
Type
as
Single
in the right menu. Select
Single
as
Edge
in the right menu.
In Edge Trigger mode, the trigger setting information is displayed on bottom right of the
screen, for example,
,
indicates that trigger type is edge, trigger
source is CH1, coupling is DC, and trigger level is 0.00mV.
Edge menu
list:
Menu
Settings
Instruction
Type
Single
Set vertical channel trigger type as single trigger.
Single
Edge
Set vertical channel single trigger type as edge trigger.
Source
CH1
CH2
Channel 1 as trigger signal.
Channel 2 as trigger signal.
Mode
Auto
Normal
Single
Acquire waveform even no trigger occurs
Acquire waveform when trigger occurs
When trigger occurs, acquire one waveform then stop
Next Page
Enter next page
Coupling
AC
DC
Block the direct current component.
Allow all component pass.
Slope
Trigger on rising edge
Trigger on falling edge
Holdoff
100 ns - 10 s, turn the
M
rotary control to set time
interval before another trigger occur.
Holdoff
Reset
Set Holdoff time as default value (100 ns).
Prev Page
Enter previous page
Trigger Level
: trigger level indicates vertical trig position of the channel, rotate trig level
rotary control to move trigger level, during setting, a dotted line displays to show trig
position, and the value of trigger level changes at the right corner, after setting, dotted line
disappears.