background image

 

 

64 

Peak 

Peak Search

 

Off 

Marker Fctn 

NdB

 

Off 

Marker Noise

 

Off 

Frequency Count

 

Off 

Marker 

Marker

 

Trace

 

Marker Table

 

Off 

Meas 

Time Spectrum

 

Off 

ACPR

 

Off 

Channel Power

 

Off 

OBW 

Off 

Pass-Fail

 

Off 

Meas Setup 

Channel Bandwidth

 

1.000000 MHz 

Channel Internal

 

2.000000 MHz 

Channel Nums

 

Power Percent

 

99.00% 

Printer 

Page Size

 

A4 

Print Language

 

Pcl 

Printer Type

 

Black/White

 

Orientation

 

Horizontal 

Copy Number

 

 

5.14.4  [Calibration 

Access  user calibration submenu including [Calibration] and 

[Factory]. 

Содержание MP700506

Страница 1: ...i MP700506 and MP700507 Handheld Spectrum Analyzer User Manual ...

Страница 2: ...10 4 6 Rear Panel Overview 12 4 7 User Interface Overview 12 4 8 Function Keys 15 4 9 Parameter Input 18 4 10 Build in Help 20 4 11 Touchscreen Controls 21 4 12 Basic Measurement 26 5 Menu Interpretation 30 5 1 FREQ bottom softkey 30 5 2 Span bottom softkey 33 5 3 AMPTD bottom softkey 34 5 4 Auto bottom softkey 38 5 5 BW bottom softkey 38 5 6 Marker bottom softkey 40 5 7 Trace hardkey 47 5 8 Detec...

Страница 3: ...key 58 5 15 File hardkey 67 5 16 Measure hardkey 69 6 Specification 73 7 Troubleshooting 81 8 Appendix 82 Appendix A Enclosure 82 Appendix B General Care and Cleaning 83 Appendix C USB Disk Requirements 83 Appendix D PC Software Requirements 83 ...

Страница 4: ... 15 Figure 4 6 Shift key and function hardkeys 17 Figure 4 7 Numeric keypad 18 Figure 4 8 The rotary control 19 Figure 4 9 Arrow keys 19 Figure 4 10 Full Span 27 Figure 4 11 Set frequency span 28 Figure 4 12 Set reference level 29 List of tables Table 4 1 AC DC adapter requirement 7 Table 5 1 Detector type comparison 49 Table 5 2 Factory Settings 62 ...

Страница 5: ...o the instrument Use Proper Overvoltage Protection Make sure that no overvoltage such as that caused by a thunderstorm can reach the product or else the operator might expose to danger of electrical shock Do not operate without covers Do not operate the instrument with covers or panels removed Avoid exposed circuit Be careful when working on exposed circuitry to avoid risk of electric shock or oth...

Страница 6: ...as filter attenuator etc as such stress may cause damages to devices and the instrument Do not mix the use of 50Ω and 75Ω connectors and or cables Do Not Overload the Input To avoid damaging the instrument the signals at input terminal must be less than 50V DC voltage components and 30 dBm 1 W AC RF components Appropriate Use of Power Meter If you are not sure of the characteristics of signal unde...

Страница 7: ...3 Handling Safety Please handle with care during transportation to avoid damages to buttons control interfaces and other parts on the panels ...

Страница 8: ...possible hazard or injury CAUTION Indicates potential damage to the instrument or other property Safety Symbols Symbols on the product The following symbols may appear on the product Refer to Manual Conforms to European Union directives This product complies with the WEEE Directive 2002 96 EC marking equipment The affixed product label indicates that you must not discard this electrical electronic...

Страница 9: ...ation This chapter lists spectrum analyzer s specification parameter Trouble Shooting This chapter helps to implement the troubleshooting and deal with after sale repair Appendix This chapter introduces accessories of spectrum analyzer and how to maintain device Convention on button and menu key format Button character bold bracket e g FREQ stands for FREQ bottom softkey Submenu words bracket e g ...

Страница 10: ...2 Check the Accessories The supplied accessories are described in the Appendix A Enclosure of this Manual Please ensure that all the listed accessories are present and undamaged if any problems are found please contact your distributor 3 Check the Complete Instrument If there is any physical damage operational fault or performance issue please contact your distributor If there is any damage to the...

Страница 11: ...st bands 2 Conductive ground mat to connect feet via ankle straps Implement both protection methods will provide a good level of anti static protection If used alone the protection will not be as reliable To ensure user s safety anti static components should offer at least 1MΩ isolation resistance Make good use of anti static technology to protect components from damage 1 Quickly ground the intern...

Страница 12: ...m RF input level of an average continuous power is 30 dBm or 50 VDC signal input Avoid connecting a signal into the analyzer that exceeds the maximum level 1 Press the power switch on the front panel 2 Self initialization takes about 30 seconds after the boot screen the spectrum analyzer will default to the scanning curve 3 After power on let the spectrum analyzer warm up for 30 minutes for stabil...

Страница 13: ...or Adjusts the screen backlight according to the environmental light This state can be specified in Display ScreenSetting Brightness Auto Power key Long push to turn on the analyzer push to turn on off the LCD display long push to turn off the analyzer Preset key Resets the analyzer to a known state This state can be specified in System PowerOn Preset Preset Arrow keys Rotary control ...

Страница 14: ...emale connector Connects to an external TTL signal The TTL signal is used to trigger the analyzer s internal sweep Shift key Press the Shift key to active it the light is on When the Shift key is active pressing a number key will execute the upper function Long press the Shift key to active it persistently To switch to the brief active status short press the Shift key Backspace key Deletes the las...

Страница 15: ...n light when external DC power is connected Red light when the battery is charging Blinking light when charge suspend or battery absent GPS BDS antenna connector SMA female type Connects an GPS BDS Antenna to offer the GPS location for the field testing Active antenna supported USB Device interface MICRO USB type Connects to a PC This USB interface supports PictBridge printer and remote control co...

Страница 16: ...Figure 4 3 Rear panel NO Name Battery compartment Air vents Tilt stand 4 7 User Interface Overview Figure below illustrates the user interface with touch screen menu keys top bar and measurement settings and results around the graph area ...

Страница 17: ...r Sweep status Set the sweep status to Single or Cont continuous Sweep Sweep Single or Sweep Cont Trigger type Set the trigger type to Auto Video Pos external positive edge Neg external negative edge Trig Source Show the source type as CW Continuous Wave or TG Tracking Generator press to turn on off the source output Source Output and Source GEN 6 16 20 14 12 21 4 18 3 5 22 19 17 15 13 2 1 8 7 27 ...

Страница 18: ...ntinuously get average of the trace to smooth it BW Average Submenu softkeys Menu items of current function can be hidden Display UI Setting Menu Hiding LAN access sign LAN access sign USB storage device status The USB icon changes to green when USB storage device is inserted Temperature or ANT Display device internal temperature If overcurrent occurs to the GPS BDS antenna red ANT will be shown S...

Страница 19: ...tector type Detector Reference level Reference level AMPTD Ref Level Amplitude Scale Type Log logarithmic or Line linear AMPTD Scale Type Amplitude Scale Display amplitude scale AMPTD Scale Div 4 8 Function Keys There are 6 function softkeys on the interface screen and 11 function hardkeys on the front panel 4 8 1 Function Softkeys There are 6 function softkeys horizontally arranged along the bott...

Страница 20: ...ns them on and off MarkerFctn Accesses the menu of special functions such as N dB bandwidth measure marker noise and frequency counting Marker Accesses the marker function softkeys that allow you to set other system parameters based on the current marker value Peak Places a marker on the highest peak and accesses the Peak functions menu 4 8 2 Function Hardkeys Pressing the Shift key followed by a ...

Страница 21: ...abel Sweep Accesses the softkeys that allow you to set the sweep time select the sweep mode of the analyzer Trig Accesses the softkeys that allow you to select the trigger mode of the analyzer Source Accesses the softkeys that allow you to set the tracking generator Demod Accesses the softkeys that allow you to set the demodulation System Sets the system parameters Accesses the calibration menu Se...

Страница 22: ...ameter value can be entered using the numeric keypad control and arrow keys 4 9 1 Numeric keypad Enters a specific value the submenu typically shows the selectable units Press the desired unit or press Enter key to complete the entry Figure 4 7 Numeric keypad 1 Number keys Numbers 0 9 are available to be used 2 Decimal point A decimal point will be inserted at the cursor position when this key is ...

Страница 23: ...nt unit for the parameter automatically 4 9 2 Rotary control Figure 4 8 The rotary control The control function During parameter editing turn the control clockwise to increase or counterclockwise to decrease the parameter values at specified steps 4 9 3 Arrow keys Figure 4 9 Arrow keys The arrow keys have following functions 1 Increase or decrease the parameter value at specific steps while editin...

Страница 24: ...e than one page of information you can read the complete information by using the arrow keys or dragging via touch screen 3 Close the current help information Press Shift key then Help key again to close help 4 Acquire the softkey help A message about how to obtain help information will be shown press the softkeys to get the corresponding help 5 Acquire the help information of any function hardkey...

Страница 25: ...rols enclosed in brackets to do the same thing Click the label to toggle between options on the top bar See User Interface Overview for details on page 12 Select a menu softkey Touch the menu softkeys in the bottom or in the right Switch menu items If there are options that can be switched in the menu you can repeatedly touch the area of the menu item to switch Press repeatedly to switch the optio...

Страница 26: ...ed the image will be saved to USB device otherwise the local memory A file name is automatically created using the current date and time stamp Double tap to capture the screen Move the active marker to the desired frequency Marker Marker turn the control When a marker is active single tap in the display area at any level to quickly move the marker to the desired frequency ...

Страница 27: ...ker to the desired frequency Set the reference level AMPTD Ref Level Swipe up or down in the display area Control the reference level Set the center or start frequency FREQ Center Freq or Start Freq Swipe left or right in the display area ...

Страница 28: ... RBW When BW RBW softkey is selected pinch and spread horizontally to change the RBW Set the VBW BW VBW When BW VBW softkey is selected pinch and spread horizontally to change the VBW Pinch and spread horizontally to change the Span or RBW VBW Set the Scale Div AMPTD Scale Div In the display area pinch and spread vertically to change the Scale Div ...

Страница 29: ...25 Pinch and spread vertically to change the Scale Div ...

Страница 30: ...the spectrum analyzer and ensure it is warmed up for 30 minutes to ensure measurement accuracy 1 Equipment connection Connect the output terminal of signal generator to the RF Input 50Ω connector on the top of the spectrum analyzer Set the parameters as follows Frequency 100 MHz Amplitude 10 dBm 2 Setting parameters 1 Press Preset on the front panel to restore the analyzer to its factory defined s...

Страница 31: ...ue the control and arrow keys can also be used to set the center frequency 3 Setting frequency span Press Span softkey input 1 and press MHz as its unit Press BW softkey set RBW to Man and input 30 and press kHz as its unit Press Shift key followed by the Detector key set the detection type to Pos Peak Figure 4 11 shows the signal at a higher resolution Please note that resolution bandwidth video ...

Страница 32: ...p right of the display area 5 Setting amplitude The reference level will be shown at the top of the display grid To get a better dynamic range the real signal peak point should be located at or near the top of display grid reference level The reference level is also the maximum value on Y axis Here we reduce to 20dB reference level to increase the dynamic range Press AMPTD softkey on the bottom an...

Страница 33: ...29 value near the top of the grid The balance between the signal peak value and noise is dynamic range Figure 4 12 Set reference level ...

Страница 34: ...nter fstop fstart and fspan denotes the center frequency the stop frequency the start frequency and the span respectively 5 1 1 Center Freq Sets the center frequency of the sweep When pressed the frequency mode is switched to Center Freq and Span in order to enter the desired parameter data Key Points The start and stop frequencies vary with the center frequency when the span is constant Changing ...

Страница 35: ...eq will increase automatically to the same value of start freq 5 1 3 Stop Freq Sets the stop frequency of the sweep When pressed the frequency mode is switched to Start Freq and Stop Freq in order to enter the desired parameter data Key Points Modifying the stop frequency changes the span and center frequency and the change of span influences other system parameters see Span You can modify this pa...

Страница 36: ...0 MHz signal you can use set both the center frequency and frequency step to 300 MHz and press the up direction key continuously to increase the center frequency to 600MHz that is secondary harmonic Press frequency steps to increase center frequency by 300MHz which reaches 900MHz Frequency Step Auto Manual shows the auto or manual mode to setting the steps When step is under manual mode press Freq...

Страница 37: ...frequencies are changed with the span automatically In manual span mode the span can be set down to 0 Hz that is zero span mode And up to the full span described in Specification When it is set to the maximum span it enters full span mode Modifying the span in non zero span mode may cause an automatic change in both CF step and RBW if they were in Auto mode and the change of RBW may influence VBW ...

Страница 38: ...e parameters signals under measurement can be displayed at an optimal view with minimum error The amplitude submenu includes Ref Level Attenuation Auto Manual Scale Div Scale Type Lin Log Ref Offset Ref Unit and Preamplifier On Off 5 3 1 Ref Level Activate reference level function and sets the maximum power or voltage for display window Key points This value is affected by a combination of maximum...

Страница 39: ...up to 40 dB You can adjust the reference level to ensure that the specified parameters meet the requirement Modifying the reference level may cause an automatic change in attenuation value But the change of attenuation value won t influence reference level You can modify this parameter using the numeric keys control or direction keys Attenuator adjustment is to make the maximum signal amplitude pa...

Страница 40: ... under internal mixer mode In general select mV as Lin amplitude scale unit Of course there would be other units for select Key points In Log scale type The Y axis denotes the logarithmic coordinates the value shown at top of the grid is the reference level and the grid size is equal to the scale value The unit of Y axis will be automatically switched into the default dBm when the scale type is ch...

Страница 41: ...nit 5 3 6 Ref Unit Sets the unit of the Y axis to dBm dBμW dBpW dBmV dBμV W or V Key points 1 dBm Choose decibel equals to 1mW as amplitude unit 2 dBμW Choose decibel equals to 1μW as amplitude unit 3 dBpW Choose decibel equals to 1pW as amplitude unit 4 dBmV Choose decibel equals to 1mV as amplitude unit 5 dBμV Choose decibel equals to 1μW as amplitude unit 6 W Choose Watts as amplitude unit 7 V ...

Страница 42: ...ncludes RBW Auto Man RBW Mode Def Con VBW Auto Man Average On Off EMI Filter 5 5 1 RBW Auto Man Adjust the resolution bandwidth ranging from 10Hz to 3MHz Use numeric key step key or control to switch resolution bandwidth The underline under Auto or Manual means Auto mode or Manual mode Press RBW Auto Manual to have Auto underlined Then the resolution bandwidth is under auto coupling mode Key point...

Страница 43: ...ver this may cause a longer sweep time Sweep Time is affected by a combination of RBW and VBW when it is in Auto mode VBW varies with RBW when it is set to Auto 5 5 4 Average On Off Trace average function It doesn t need narrow VBW to implement smooth trace This function set the wave detector to acquisition mode continuously get average of the trace to smooth it 5 5 5 EMI Filter Pop out the menu f...

Страница 44: ...or time when any marker type menu is active so as to view the readouts of different points on the trace 5 6 1 1 Marker 1 2 3 4 5 Selects one marker the default is Marker1 and place the marker at the center of the trace If a delta marker is activated this softkey changes to the menu under the Delta function If there is already a marker this command will not produce any operation If there are alread...

Страница 45: ...5 In the trace measurement the marker used to activate the traces 5 6 1 3 Normal One of the marker types which is used to measure the values of X Frequency or Time or Y Amplitude at certain point of the trace When selected a marker will appear with its marker number such as 1 on the trace Key points If no active marker exists currently a sing frequency marker will be activated automatically at the...

Страница 46: ...hile the Delta Marker is active You can use the control touch screen or the numeric keys to move the delta marker The frequency or time and amplitude differences between the two markers will show in the active area and in the upper right corner of the display The displayed amplitude difference is expressed in dB or is the linear unit in terms of the corresponding scale If Delta has been activated ...

Страница 47: ... 6 2 2 NdB On Off Enables the N dB bandwidth measurement or sets the value of N The N dB bandwidth denotes the frequency difference between points that are located on both sides of the current marker while the amplitude falls off N 0 or rises N 0 N dB separately Key points When the measurement starts the analyzer will search the two points which are located at both sides of the current point and a...

Страница 48: ...er function This function is invalid when the tracking generator is activated The count value is displayed in the upper right corner of the screen 2 BW Sets the frequency counter resolution The allowed values are 1 Hz 10 Hz 100 Hz and 1 kHz Changing the counter resolution can change the counter accuracy The higher the resolution the higher the counting accuracy 5 6 3 Marker Key access Marker botto...

Страница 49: ...e analyzer to the frequency of the current marker In Normal marker mode the start frequency will be set to the frequency of the current marker In Delta marker mode the start frequency will be set to the frequency at which the Delta Marker is located The function is not available in zero span mode 5 6 3 4 Mkr Stop Sets the stop frequency of the analyzer to the frequency of the current marker In Nor...

Страница 50: ...ak Key access Marker bottom softkey Peak right softkey Accesses the Peak setting menu Key Points The spurious signal at the zero frequency caused by LO feed through is ignored 5 6 4 1 Max Search Place a frequency marker at the highest point of the trace and display the frequency and amplitude of the marker in the upper right corner of the screen Max search does not change the activated function 5 ...

Страница 51: ...oint 5 6 4 7 Cont Max On Off Enables or disables the Cont Max search the default is Off When enabled the system will always execute a peak search automatically after each sweep in order to track the signal under measurement 5 7 Trace hardkey Key access Press Shift key then Trace hardkey 7 in numeric keypad As the sweep signal is displayed as a trace on the screen you can set parameters about the t...

Страница 52: ...t if a new minimum level is detected in successive sweeps 5 7 5 Blank Clear the trace on screen The selected trace register stores the trace data and will not be updated as the analyzer sweeps 5 7 6 View Holds and displays current trace for observation The trace register is not updated as the analyzer sweeps 5 7 7 Operations Enter the trace math submenu 1 1 2 Exchange the trace register 1 data wit...

Страница 53: ...rmation associated with a larger subrange That is several samples may fall on one pixel Which of the samples will be represented by the pixel depends on the selected detector type The submenu includes Pos Peak Neg Peak Normal Sample RMS Avg Voltage Avg Quasi Peak Key points Selects an appropriate type according to the application in order to ensure the accuracy of the measurement for your applicat...

Страница 54: ...tage Average detector averages the linear voltage data of the envelope signal measured during the bucket interval It is useful for observing rise and fall behavior of AM or pulse modulated signals Quasi Peak Quasi peak detector is a weighted form of peak detection The measured value drops as the repetition rate of the measured signal decreases It is used in EMI testing 5 8 1 Pos Peak Searches the ...

Страница 55: ... mode is available when EMI filter is turned on The quasi peak detector is a peak detector that is weighted by the duration and repetition rate of the signal as specified by the CISPR 16 1 1 standard Quasi peak detection is characterized by a fast charge time and slow decay time 5 9 Display hardkey Key access Press Shift key then Display hardkey 9 in numeric keypad Controls the screen display of t...

Страница 56: ...tal reference line is activated on the screen 5 9 3 2 Ampt Graticule On Off Turn on or Off amplitude scale function 5 9 3 3 Grid On Off It s the grid lines displaying and hiding menu When the grid display line is on pressing Grid On Off again will turn it off 5 9 3 4 Label On Off Defines the content displayed or hidden in the comments that appear in the display grid area 5 9 3 5 Menu Hiding On Off...

Страница 57: ...e time 1 to 60 minutes Press the power key to re activate the LCD display after the LCD display sleeping mode has been triggered 5 10 Sweep hardkey Key access Press Shift key then Sweep hardkey 4 in numeric keypad Sets parameters about the Sweep time and mode including Sweep Time Auto Manual Sweep Single Sweep Cont Sweep Points 5 10 1 Sweep Time Auto Man Sets the time interval for the analyzer to ...

Страница 58: ...DC Analog to Digital Converter That is the more points used the longer the sweep will be Changing the points would also influence other system parameters thus the instrument restarts the sweep and measurement cycle You can modify this parameter using the numeric keys control or direction keys 5 11 Trig hardkey Key access Press Shift key then Trig hardkey 5 in numeric keypad Sets the trigger mode a...

Страница 59: ...urce hardkey 6 in numeric keypad When the Source is turned on an independent signal or a signal with the same frequency of the current sweep signal will be output from the GEN Output 50Ω terminal on the top panel Press the key to access the submenu includes Output On Off Source GEN CW TG Output Level Output FREQ OF Step Network Meas The source is turned off in the power on and reset states 5 12 1 ...

Страница 60: ...relative values after normalized expressed in dB When the network measurement function is off the measurement shows the spectrum measurement result expressed in dBm 1 Network Meas On Off Turn on or off the tracking source network measurement function Tracking source network measurement function mainly for amplitude and frequency characteristics measurement RF output and spectrum measurement is ful...

Страница 61: ...ce network measurements After connecting the instrument s RF output to the RF input press the normalized soft menu and the display shows a straight line on the 0dB scale 5 13 Demod hardkey Key access Press Shift key then Demod hardkey 1 in numeric keypad Enter the demodulation settings the spectrum analyzer supports audio demodulation and AM FM analog demodulation 5 13 1 Audio Demod Enter Audio De...

Страница 62: ...on On or Off 2 Carrier Freq Set the carrier frequency of the AM modulation signal 3 IF BW Auto Man Set the demodulation bandwidth to auto or manual mode 5 13 2 2 FM Enter FM demodulation submenu 1 FM On Off Turn FM demodulation On or Off 2 Carrier Freq Set the carrier frequency of the FM modulation signal 3 IF BW Auto Man Set the demodulation bandwidth to auto or manual mode 5 14 System hardkey Ke...

Страница 63: ... root directory of the USB memory device and copy the firmware file onto this folder 2 Insert the USB memory device into the top panel USB connector on your instrument Press Shift key then System hardkey Press System and press Firmware Update to execute firmware update 3 The analyzer will perform the update process The upgrade procedure will take approximately 30 seconds During the update process ...

Страница 64: ... address subnet mask and gateway to the analyzer on the basis of the current network status 5 14 2 2 Control Access the submenu for touch control and other control settings 1 TouchControl Access the submenu for touch control Touch Control Turn on off the touch control Align X When dragging up or down in the display area the reference level is changed to align to the grid of X axis Align Y When dra...

Страница 65: ...screen menu language 5 14 2 5 Date Time Sets the date time and display format of the analyzer Date Time On Off Turn on or off date time display Format Selects the display format Date and time can be displayed as YMD HMS or HMS YMD Date Set Set the date for the analyzer Use the numeric keys to enter the date The format is YYYMMDD E g June 22th 2012 should be entered as 20120622 Time Set Set the tim...

Страница 66: ...y press Save Recall and select the User State menu item Last Restores the analyzer to the user setting at last shutdown Table 5 2 Factory Settings Parameter Value Frequency Center Frequency MP700506 800 009000 MHz MP700507 1 800009000 GHz Start Frequency 9 000 kHz Stop Frequency MP700506 1 600009000 GHz MP700507 3 600009000 GHz Frequency Step MP700506 Auto 160 000000 MHz MP700507 Auto 360 000000 M...

Страница 67: ... VBW Auto 3 MHz Trace Average Off Detector Detector Type Pos Peak Sweep Sweep Time Auto 20 000 ms Sweep Mode Continuous Sweep Source Source TG Off Network Meas Off Trace Trace 1 Trace Type Clear Write Trace 1 Operations 1 2 Display Full Screen Off Window Zoom Off Display Line Off Amplitude Graticule Off Grid On Label On Trig Trigger Type Auto Run ...

Страница 68: ...f ACPR Off Channel Power Off OBW Off Pass Fail Off Meas Setup Channel Bandwidth 1 000000 MHz Channel Internal 2 000000 MHz Channel Nums 3 Power Percent 99 00 Printer Page Size A4 Print Language Pcl Printer Type Black White Orientation Horizontal Copy Number 1 5 14 4 Calibration Access user calibration submenu including Calibration and Factory ...

Страница 69: ...tatus 5 14 5 Printer Access the relative soft menu for printing 5 14 5 1 Page Size Select the paper size to be printed including A4 A3 B5 C5 Letter 5 14 5 2 Language Pcl Esc Select the print language optional for Pcl and Esc 5 14 5 3 Printer Type Blak Color Select the printer type to gray or color 5 14 5 4 Orientation Hor Ver Select the print orientation to horizontal or vertical 5 14 5 5 Number S...

Страница 70: ... trace data file format is CSV The file name is automatically created using the current date and time When Save to USB is selected the trace file will be saved in a folder named spectrum created automatically in USB disk 5 14 6 3 User State Save the current system setting as a user self defined setting to local memory The user setting file can be used as the power on setting or preset setting 5 14...

Страница 71: ...e 6 Load Recall the selected file 5 15 File hardkey Key access Press Shift key then File hardkey 3 key in numeric keypad Access the file management menu 5 15 1 Refresh Update the file list in the directory 5 15 2 Type Select the file type you wish to view under the directory includes screen image trace data user states or display all Screens png A screen file is the captured graphic of the screen ...

Страница 72: ...he file list 5 15 7 2 Delete Press Delete Select to delete the selected file Press Delete Page to delete the files in current page Press Delete All to delete all the files 5 15 7 3 Export When a USB disk is inserted you can export the files in the local memory to the USB disk Press Select to export the selected file to USB disk Press Page to export the files in current page to USB disk Press All t...

Страница 73: ...spectrum adjacent channel power measurement channel power measurement occupied bandwidth Pass Fail measurement menu 5 16 1 Measure off You can directly close the currently running measurement function you can also choose to close the measurement menu 5 16 2 Time Spec On Off Turn on time spectrum measure mode 5 16 3 ACPR On Off Turn on or off the adjacent channel power measurement Press Meas Setup ...

Страница 74: ... the absolute value of the main channel power 5 16 5 OBW On Off Turn on or off the occupied bandwidth measurement Press Meas Setup to enter the parameter setting submenu for occupying the bandwidth measurement Occupied Bandwidth is a measure of the bandwidth occupied by the transmitter signal can be measured from the total power ratio within the in band power span with a default value of 99 the us...

Страница 75: ... frequencies for scanning line for editing 6 Window Sweep On Off Turns window sweep on or off When the window sweep is on only the window formed by the intersection of the amplitude line and the frequency line is scanned The peripheral stops scanning the full frequency is scanned when it is closed 5 16 6 2 Limit Meas Enter the submenu of the limit measurement mode 1 Limit Meas On Off Turn On or Of...

Страница 76: ... on the whole superimposed on a degree so that it can move up or down easy to measure Does not affect the amplitude setting of the spectrum analyzer 5 UpLine Edit Upper line editing is used to edit the control line above the trace depending on the trace 6 LowLine Edit Lower line editing is used to edit the control line above the trace depending on the trace 5 16 7 Meas Setup Measurement setting me...

Страница 77: ...ions of the spectrum analyzer Unless otherwise stated the technical specifications apply to the following conditions The instrument has been preheated for 30 minutes before use The instrument is in the calibration cycle and has been self calibrated Typical and nominal for this product are defined as follows Typical Refers to the performance of the product under certain conditions Nominal Refers to...

Страница 78: ...0 kHz 80 dBc Hz Typical 100 kHz 100 dBc Hz Typical 1 MHz 107 dBc Hz Typical Bandwidth Resolution Bandwidth 10 Hz to 500 kHz 1 10 steps by sequence 1 MHz 3 MHz RBW Accuracy 5 typical RBW 1 MHz Resolution Filter Shape Factor 60 dB 3 dB 5 1 typical digital and close to Gaussian shape Video Bandwidth VBW 10 Hz to 3 MHz Amplitude and level Amplitude measurement range DANL to 20 dBm 1 MHz to 3 6 GHz Pre...

Страница 79: ...38 dBm Typical 128 dBm MP700507 1 GHz to 3 6 GHz Preamp On 1 MHz to 1 GHz 160 dBm Typical 150 dBm MP700506 1 GHz to 1 6 GHz 158 dBm Typical 148 dBm MP700507 1 GHz to 3 6 GHz Frequency response 20 to 30 30 to 70 relative humidity input attenuation 21 dB reference frequency 50 MHz Preamp Off fc 100K 0 8 dB 0 4 dB Typical Preamp On fc 1 MHz 0 9 dB 0 5 dB Typical Difference and Accuracy RBW Switch Dif...

Страница 80: ...dBm to 50 dBm 1 5 dB VSWR Input 10 dB RF attenuation 1MHz to 3 6 GHz 1 5 Nominal Distortion and spurious response Second harmonic distortion fc 50 MHz Preamp off signal input 10 dBm 0 dB RF attenuation 20 to 30 65 dBc Third order intermodulation fc 50 MHz Input double tone level 20 dBm frequency interval 100 kHz input attenuation 0 dB preamplifier off 20 to 30 10 dBm 1 dB Gain Compression fc 50 MH...

Страница 81: ...GHz Tracking Source Output 35 MHz to 3 6 GHz Signal generator Output power level range MP700506 30 dBm to 0 dBm MP700507 40 dBm to 0 dBm Output power level resolution 1 dB Output flatness 3 dB Maximum safe reverse level Average total power 30 dBm DC 50 VDC Demodulation Audio Demod ulation Frequency Range MP700506 100 kHz to 1 6 GHz MP700507 100 kHz to 3 6 GHz Demodulati on Type FM AM AM Measur eme...

Страница 82: ... Modulation Rate Accuracy 1 Hz nominal Modulation rate 1 kHz 0 1 modulation rate nominal Modulation rate 1 kHz Deviation 20 Hz to 200 kHz Deviation Accuracy 4 nominal Frequency Counter Resolution 1Hz 10Hz 100Hz 1kHz Accuracy frequency indication frequency reference accuracy counter resolution RF Input Impedance 50 Ω typical Connector N Type Female Tracking Generator Output Impedance 50 Ω typical C...

Страница 83: ...tocol USB 2 0 Host End USB Device Connector Mini Protocol 2 0 Version Display Type TFT LCD Resolution 1024 768 Size 8 inches Color 65536 Remote Control USB USB TMC LAN 10 100 Base RJ 45 Mass Memory Mass Memory Internal Storage 256M Bytes Temperature Operating Temperature Range 0 to 40 Storage Temperature Range 20 to 60 Appearance ...

Страница 84: ...80 Dimensions 265 mm W 190 mm H 58 mm D Weight Approx 2 5 kg without package ...

Страница 85: ... connector If there is still no signal display there may be a problem with the spectrum analyzer hardware circuit Please contact your supplier for service 3 Wrong measurement results or poor signal frequency precision If the display contents shake a lot or the frequency readout exceeds the error range during measurements check if the signal source is stable If so check if spectrum analyzer referen...

Страница 86: ...d Accessories Power Cord CD Rom Quick Guide USB Cable AC DC Adapter GPS Antenna Metal Case Options N N Cable N SMA Cable SMA SMA Cable SMA Adaptor N SMA Adaptor Near Field Probe includes Four near field probes N SMA adapter SMA SMA cable Frequency range 30 MHz 3 GHz Soft Carrying Case ...

Страница 87: ...lectric shock or electrical short circuit resulting from moisture Appendix C USB Disk Requirements USB disk requirements Max capacity 4G NTFS file system is not supported If the USB disk doesn t work properly format your USB disk and then try again Appendix D PC Software Requirements The PC software support Windows 10 Windows 8 Windows 7 Windows Vista Windows 2000 The PC software does not support ...

Отзывы: