51
displays at the corresponding pixel. This mode is usually used for
video averaging and noise frequency Maker.
5.8.5 [RMS Avg]
Set the detector to the RMS Average detector mode. This mode
calculates the RMS average power of all the samples in the
sample bucket.
5.8.6 [Voltage Avg]
Set the detector to the Voltage Average detector mode. This mode
calculates the average voltage of all the samples in the sample
bucket.
5.8.7 [Quasi-Peak]
Set the detector to the Quasi-Peak detector mode. This mode is
available when EMI filter is turned on. The quasi-peak detector is a
peak detector that is weighted by the duration and repetition rate of
the signal, as specified by the CISPR 16-1-1 standard. Quasi-peak
detection is characterized by a fast charge time and slow decay
time.
5.9
【
Display
】
hardkey
Key access:
Press
Shift key, then
Display
hardkey (
9
in
numeric keypad
).
Controls the screen display of the analyzer, such as full screen,
setting the on or off for window zoom, display line, amplitude scale,
grid and label.
5.9.1 [Full Screen]
Set to full-screen display graphical interface, press any key to exit.
5.9.2 [Zoom On Off]
In multi-window display mode, press this button to zoom in on the