![MULTI MEASURING INSTRUMENTS MLIT-1 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/multi-measuring-instruments/mlit-1/mlit-1_instruction-manual_1853934005.webp)
3. SPECIFICATIONS
1) INPUT PART
(1) VOLTAGE INPUT (PHASE VOLTAGE)
①
Input Method : Direct Input by Test Leads
②
Input Impedance : more than 1M
Ω
③
Input Voltage Range : AC50V
~
AC500V
④
Input Frequency : 50Hz or 60Hz by Switch
⑤
Minimum Resolution : 0.1V
(2) CURRENT INPUT (GROUNDING LINE CURRENT)
①
Input Method : Split-Core Type ZCT
2
) CURRENT DETECTION PART
①
Inside Diameter of CT :
Φ
30mm
②
Method : Split-Core Type ZCT
③
Structure : Manual Slide Method
④
Withstanding Voltage : AC2000V, 1 Minute
⑤
Length of Cable : 2000mm
⑥
Dimension & Weight : 29.6 x 90.2 x 70.7mm, Approx. 200g
3
) MEASUREING PART
①
Measuring Function : AC Voltage, Line Current, Leakage Current,
Resistive Leakage Current (Ior), Insulation Resistance
②
Measuring Method : Clamp CT Method
③
Measuring Range : Leakage & Line Current (Auto-range)
0
~
200
μ
A / 2mA / 20mA / 200mA
AC Voltage : 0
~
500.0V (1 range)
Insulation Resistance : Calculation by Current & Voltage
④
Input Frequency : 45
~
65Hz (by Frequency Change Switch)
⑤
Minimum Resolution : Leakage, Line, Ior Current : 0.1
μ
A
AC Voltage : 0.1V
⑥
AC Detection Method : Average Sensing RMS Conversion
⑦
A/D Conversion : Successive Comparison Method
⑧
Display : Max. 1999 Count LCD with Annunciators
- 4 -